THE TIME EXPANSION SYSTEM  

 
This technique is similar to recording a sound on a tape recorder and then replaying the tape at a 
reduced speed. However, in the D240x the signal is not stored on a magnetic tape but in a digital 
memory. The time expansion principle is unique in the sense that it is the only ultrasound conversion 
system that retains virtually all characteristics of the original signal, so time expanded signals are well 
suited for many forms of advanced signal analysis (e.g. with our sound analysis software package 
BatSound). A time expansion factor of either 10 or 20 can be selected with the switch located 
immediately to the left of the relay output connector. The switch is accessible through the opening for 
the relay connector. Using a small screwdriver or similar object is recommended to alter the switch 
setting. The switch was given this sheltered position on purpose, to avoid inadvertently changing its 
setting. The factory default setting is 10 (switch pointing towards the microphone).    
 
There are two basic operation modes of the time expansion system, manual and automatic. The 
operation mode is selected with the TRIG switch. In the manual mode, the MANUAL START/STOP 
switch is used to start and stop storing the signal, while in the automatic mode, the detector captures 
the signal itself when it exceeds an adjustable threshold level. 
 
The maximum storage time in the time expansion system is 3.4 seconds. The actual memory size is 
selected with the MEMory SIZE switch. The alternatives are: 3.4 sec, 1.7 sec and “MIN”. The default 
factory value for “MIN”is 0.1 sec. On request, it is however also possible to obtain a “MIN” storage 
time of 0.05 sec or 0.8 sec.  
 
In order to obtain the highest sound quality on time expanded sounds, headphones should be used to 
monitor the sounds. Listening to the heterodyne system through the speaker, while sampling a signal 
to the time expansion system will cause the high frequency parts  of the speaker sound to enter the 
microphone and to be mixed with the desired ultrasonic signal. 
 
When the D240X is switched on, the memory of the detector is filled with a random signal which may 
sound like white noise. After switching the detector on, please allow at least the chosen memory time 
(e.g. 3.4 seconds) to elapse before you interrupt the recording by pressing the MANUAL 
START/STOP button (or, in the AUTO mode, the recording is interrupted by an incoming sound) and 
this random signal will be replaced with a “real” signal. If you change the memory time, please first 
make a new recording and allow the entire memory to be filled, as described above. 
 
 

The manual triggering mode 

 
To select the manual triggering mode, set the TRIG switch to the MAN position. 
 
The MANUAL START/STOP switch is used to start and stop the sampling into the digital memory of 
the time expansion system in the manual operation mode. The red LED below the frequency display 
indicates the recording status (light on = ”recording”, light off = ”replaying”). When the detector is in 
the ”record” mode, samples are continuously stored into the memory (when the memory is full the 
samples that were first stored are overwritten, so at each time, the memory contains the last 3.4/1.7/0.1 
seconds of the signal). Pushing the push-button to switch to the ”replay” mode, will cause the content 
of the memory to be replayed in an endless loop fashion. This is done at 1/10 (1/20) of the original 
rate, so one sequence lasts for 34/17/1 (68/34/2) seconds. 
 
 

The automatic triggering mode 

 
To select the automatic triggering mode, set the TRIG switch to the AUTO position. In this mode, the 
detector will capture the signal itself and store it in the time expansion memory.  
 

Содержание D240x

Страница 1: ...ULTRASOUND DETECTOR D240x OPERATING INSTRUCTIONS www batsound com Phone 46 1830 3880 Fax 46 1830 3840 e mail info batsound com...

Страница 2: ...ckwise to switch the detector on The LCD display now shows the tuned frequency heterodyne system only and the LCD backlight is turned on The LCD backlight also serves as a battery condition indicator...

Страница 3: ...nge from appr 25 to 35 kHz will be transformed This is illustrated in the figure below Let s assume an ultrasonic signal with constant frequency is emitted When the frequency control of the D240x is t...

Страница 4: ...ime expanded sounds headphones should be used to monitor the sounds Listening to the heterodyne system through the speaker while sampling a signal to the time expansion system will cause the high freq...

Страница 5: ...signals between approximately 25 and 35 kHz This is useful to reduce the influence from background noise or to make the detector capture only certain species It should be noted that even if the detect...

Страница 6: ...nal output 3 signal 4 ground 5 polarizing hole for the connector NOTE If you wish to connect equipment other than Pettersson Elektronik s to this connector please first contact us for instructions Ple...

Страница 7: ...PHONES output connections Technical specifications Type Heterodyne and time expansion x10 or x20 Frequency range 10 120 kHz min Display accuracy 0 15 kHz Bandwidth het 8 kHz 4 kHz 6 dB Sampling frequ...

Отзывы: