8
Store Smart Check in a safe and accessible location. See
d
isconnecTion
and
s
Torage
instructions.
Smart Setup Instructions, Continued
Your cushion is now set up.
(Next: See
C
heCk
I
nstruCtIons
.)
When you see the flashing green
light, close the inflation valve. The
green light means that Setup is
complete.
5b.
Press (but do not hold) the Setup Button.
The red down arrow
light will now
flash
.
Open the inflation valve, release air (approximately 30
to 60 seconds)*, and wait for the flashing green light.
If the red down arrow
light stops flashing before you
have opened the valve, go back to 5a.
*To ensure an accurate reading:
Remain still and avoid
restricting air flow while you release air from the cushion.
5a.
Press the Check
Button.
When you see the red down arrow
light,
proceed to step 5b.
NOTE:
If you see the red up arrow
light
when you press the Check Button:
1) Make sure the quick disconnect is securely
attached, and 2) Open the inflation valve,
add more air, close the inflation valve, and
press the Check
Button again.
Repeat until
you see the red down arrow light.
5a.
5b.