Permobil ROHO QUADTRO SELECT Series Скачать руководство пользователя страница 2

QUADTRO SELECTAdjustment Guide Rev. 4/19 Print 4/19

Please refer to the Operation Manual for other safety 
information regarding the QUADTRO SELECT cushion.

NOTE: Before adjusting QUADTRO SELECT for an asymmetry, you MUST perform a hand check to adjust the air inflation level to

 

proper cushion inflation setting 

as detailed on the 

reverse side of this document. 

Adjusting a ROHO

®

 QUADTRO SELECT

®

 

cushion for positioning

1. Make sure the ISOFLO Memory Control is open

(press green knob in). FIGURE 1

2. Have the individual sit on the cushion in the most

upright and midline posture that they can achieve.
FIGURE 9

3. Close the ISOFLO (press red knob in). This will

help hold the pelvis in the best posture that is
achievable by that individual and prevent further
progression into asymmetry. FIGURE 3

4. Perform a hand check and make sure there is a

minimum of ½ - 1” of air between the lowest ischial
tuberosity and the seat surface. FIGURE 8 & 10

½"-1"

½"-1"

To correct a reducible (flexible) 

pelvic obliquity (lateral asymmetry):

1. Make sure the ISOFLO Memory Control is open

(press green knob in). FIGURE 1

2. Make sure the pelvis is positioned at the back of

the cushion. FIGURE 2

3. Position the pelvis in a neutral and upright position

by pressing down on the higher side of the pelvis
until the pelvis is level. FIGURE 6
This will force more air into the quadrant under
the lower side of the pelvis to help prevent it from
tilting to that side.  FIGURE 7
Close the ISOFLO (press red knob in), then release
your pressure on the pelvis. FIGURE 3

4. Perform a hand check and make sure there is

a minimum of ½ - 1” of air between the ischial
tuberosity and the seat surface. FIGURE 8

To correct a reducible (flexible) posterior 

pelvic tilt (anterior-posterior asymmetry):

1. Make sure the ISOFLO Memory Control is open

(press green knob in). FIGURE 1

2. Make sure the pelvis is located at the back of the 

cushion, in a neutral and upright position. FIGURE 2

3. Use your hands to create downward pressure on

both sides of the pelvis to force air to the front
quadrants and create a contour for the pelvis that
will help discourage forward migration; close the
ISOFLO (press red knob in), and then release the
pressure on the pelvis. FIGURE 3

4. Perform a hand check and make sure there is

a minimum of ½ - 1” of air between the ischial
tuberosities and the seat surface. FIGURE 4

Open ISOFLO - Push green 

knob in

Close ISOFLO -

 

Push red 

knob in

www.permobil.c

a

To accommodate a non-reducible (fixed) 

postural asymmetry:

Figure 1

Figure 2

Figure 3

Figure 4

Figure 8

Figure 7

Figure 5

Figure 6

Figure 9

Figure 10