Page: 6/15
Programming manual
C
ompact
J
oystick
A
dvanced R-net
Document Number: 205314-UK-0 Edition 2 2009-10 (0033-8009a-0100a)
4.2.1.2 Group 1 ... 4 Drive Axis
Sets which axis is driven when the Group 1 … 4 function is Drive Axis
1 ... 32
This number represents the axis that will be operated
when you have selected
‘Drive Axis’
in
‘group’
.
4.2.2 Grab Focus: Default – Local On
Sets how focus can be grabbed.
Standard:
The CJA will function similar as a standard JSM from
PGDT
(refer
to
R-Net Control System Technical Manual
–
SK77981
– from PGDT).
Local On:
If the local power button/switch is operated to start-up the
system, the joystick module will be the active input
device.
Local On/JS Movement:
Additional to the Local On the focus can be grabbed by
any movement of the joystick *.
Local On/JS Double Reverse:
Additional to the Local On the focus can be grabbed by
moving the joystick twice backwards *.
* See also Grab Focus Time