Local output mode:
The way to operate you actuators, lights or horn depends on how the CJA is
programmed. Up to 16 different outputs (e.g. recline, left indicator, horn, etc.) can
be programmed. The outputs are located in 4 groups and each group can contains
up to 4 outputs.
If you enter the local output mode you need first to select a group. The group is
indicated by a flashing led.
Group 1
Group 2
Group 3
Group 4
After you selected a group you can select the desired output. The output is
indicated by a LED.
Output 1
Output 2
Output 3
Output 4
If your wheelchair is setup to use only up to 4 outputs in 1 group, the group
selection is not needed.
2.5 Diagnostic Screen or Acoustic Feedback
When the control system safety circuits have operated and the control system has been
prevented from moving the wheelchair, a diagnostics code will be displayed.
This indicates a system trip, i.e. the R-net has detected a problem somewhere in the
wheelchair’s electrical system.
All red LED’s will flash periodically alternated with a green flash of the battery/service
indicator. The number of red flashes is an indication of the fault. We refer to the State
Indication Chart. Se Chapter 3.11 in this manual
If the error is not critical, for example the ICS detects a broken light, then drive will still be