84 . TGA 8000 Installation and Hardware Guide
Needle Replacement/Adjustment Procedure
The needle should be replaced if it is old or contaminated. If a different size pan is used, you
may need to adjust the needle in the stem in order to maintain the 5-mm space between the
needle's tip and the top of the pan.
1.
Remove the Accupik to replace the needle.
2.
Using a screwdriver, loosen the screw that is attached to the metal shaft. Pull the old
needle out and insert a replacement needle (P/N N5376668).
3.
Replace and tighten the screw with the screwdriver.
4.
Replace the Accupik onto the TGA 8000.
5.
Check the alignment and adjust the Accupik as necessary.