Control Panel

1. Run signal light

2. Time signal light

3. Display Screen

4. Frequency signal light

5. Control Knob

6. Start/Stop button

1.

2.

3.

4.

5.

6.

RUN

START

STOP

TIME

FREQ

• RUN

 Signal light (green) - lit when the Bead Ruptor

TM

 96 Homogenizer is running.

• TIME

 Signal light (yellow) - lit when the Bead Ruptor

TM

 96 Homogenizer timer  

 

  selection is active.

• FREQUENCY

 Signal light (yellow) - lit when the Bead Ruptor

TM

 96 Homogenizer  

 

  frequency selection is active. Frequency is defined as number of wave cycles per  

 

  second (1/s). Measured in Hz.

• CONTROL KNOB

 (rotate / push) - rotate right (+) Or left (-) to change the Time

  or Frequency setting values. Push the Control Knob to switch between Time and

  Frequency settings.

• START/STOP

 button - press to start or stop the cycle.

WARNING:

 Only start the Bead Ruptor

TM

 96 Homogenizer with the lid in the closed 

position. The display will read “LID” if the lid is open when attempting to start. 

Always start and stop the Bead Ruptor

TM

 96 with the Start/Stop button.

Содержание OMNI International Bead Ruptor 96

Страница 1: ...Bead RuptorTM 96 Well Plate Homogenizer...

Страница 2: ...EXCEED THE CONTRACT PRICE FOR THE GOODS FOR WHICH LIABILITY IS CLAIMED This product has been engineered for safety however basic safety precautions and common sense must always be demonstrated when u...

Страница 3: ...200W Fuses 2 x 2AT 250V 230V 2 x 5AT 250V 115V Vibrational frequency regulation Digital from 3 to 30 Hz 180 1800 min 1 in 0 1 Hz steps Timer 5 sec 99 min timer HOLD function in 1 sec steps below 10 mi...

Страница 4: ...e Bead RuptorTM 96 Homogenizer frequency selection is active Frequency is defined as number of wave cycles per second 1 s Measured in Hz CONTROL KNOB rotate push rotate right Or left to change the Tim...

Страница 5: ...A A B B C C A Centering Points B Locking Pin C Knob Milling Jars O Rings Jar Wrenches 96 Well Plate And 2D Barcode 96 Well Storage Tube Adapters Milling Jar and Rack Holders...

Страница 6: ...t 90 to unlock and turn the knob C counter clockwise to loosen If locking pin is difficult to release turn knob C slightly clockwise to release locking pin tension WARNING Always include an O Ring whe...

Страница 7: ...r clockwise all the way to maximize clamping range Turn locking pin 90 and allow it to reengage in locked position Place the well plate adapter into the milling jar arm and seat it into centering poin...

Страница 8: ...when working with liquid nitrogen Liquid nitrogen has a temperature of 196 C and will cause severe burns and frostbite with skin or eye contact Milling Jars Order Number Stainless Steel 25 mL set of t...

Страница 9: ...d push it until it locks 1 Fuse Compartment 2 Power Cord Port The following fuses are required for the Bead RuptorTM 96 Homogenizer 2 x 2AT 250V 230V 2 x 5AT 250V 115V Unplug power cord from the power...

Страница 10: ...PN 03 296 Rev G 935 Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com...

Отзывы: