Introduction . 15
Sample Injection/Wash Port
The Sample Injection/Wash Port (Figure 6 and Figure 7) consists of a waste transfer and
port. After completion of a liquid handling function, the excess sample is discarded in the
waste port. The probe is then moved to the wash port for cleaning. Washing is accomplished
by pumping wash fluid through the probe. The fluid then flows around the outside of the
probe and overflows into the waste port. The waste fluid then travels by gravity out of the
waste tubing to a waste reservoir located below the instrument.
Figure 6 Sample injection and wash ports (side view)
Figure 7 Sample injection and wash ports (top view)
The transfer port is used to transfer the sample to the flowcell located in the sample
compartment of the spectrometer. The transfer port consists of a nylon fitting into which the
sample probe is lowered.
Sampling Probe/Needle
The sample is drawn through the stainless steel needle and delivered to the flowcell via the
transfer port. The needle is cleaned in the wash port: the solvent flows down inside the
needle, and then is forced up around the outside and to waste.
Waste exit port
Sample injection &
wash ports
Wash station
Waste position
Injection port
Interconnect
tubing port
Bolt
v1
v2
v3
Содержание OilExpress System
Страница 1: ...OilExpress System User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 23: ...Warnings and Safety Information ...
Страница 36: ...36 OilExpress System User s Guide ...
Страница 37: ...Unpacking and Installation ...
Страница 69: ...Background to Used Oil Analysis ...
Страница 75: ...Routine Maintenance and Troubleshooting ...
Страница 96: ...96 OilExpress System User s Guide ...
Страница 97: ...Appendices ...
Страница 102: ...102 OilExpress System User s Guide ...