09931262E
8
b.
Experiment: SBW is set 2.0 nm automatically. Setup other experiment parameters.
c.
Lamp: Single Front is automatically selected. If not, select Single Front and setup other
experiment parameters.
d.
Setup the Internal Reference, Scan Setup. For more information, see the
Lambda 365 UV
Express Software Users Guide, the section on Scan Setup.
5.
Select Apply and OK after setting up the parameters.
6.
Dip the probe in the blank solution and select the Blank icon.
NOTE:
Before measurement, make sure that that bubble does not exist at the mirror side of the probe.
NOTE:
When the sample is measured, any light from outside of the bottled should be blocked as it affects the
measurement result.
7.
Clean the probe and dip it in the sample solution. Select the Sample icon.
8.
Input the sample name and select OK.
9.
The spectrum and result will be displayed. Save or print the results as desired.