Advanced Maintenance . 75
2.
Slacken the two terminal screws (Figure 42) securing the wires leading to the source.
Disconnect the wires from the connectors.
For the Frontier FT-IR spectrometer, the connecting wires for the MIR source are white.
For the Frontier FT-NIR spectrometer, the connecting wires for the NIR source are red.
Figure 42 Terminal screws (MIR source shown)
3.
By grasping the round metal source body, NOT the wires, remove the source from its
housing by pulling, noting its installed position.
4.
Install the new source in its housing, by pushing in until it clicks into place.
5.
Reconnect the two wires to the terminals and secure by tightening the two terminal
screws.
6.
Close and secure the main cover.
Refer to page 72, steps 10 to 15.
Allow the source temperature to stabilize for approximately 2 hours before using the
instrument.
Source
Source housing
Terminal screw
Terminal screw
Top of connector block
Содержание Frontier FT-IR
Страница 1: ...FRONTIER IR SINGLE RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Frontier IR Single range Spectrometers User s Guide ...
Страница 11: ...Warnings and Safety Information ...
Страница 23: ...An Overview of the Frontier IR Single Range Spectrometers ...
Страница 32: ...32 Frontier IR Single range Spectrometers User s Guide ...
Страница 33: ...Unpacking and Installation ...
Страница 43: ...Using the Spectrometer with Spectrum ...
Страница 49: ...Using the Spectrometer with Spectrum 49 Figure 20 Atmospheric suppression ...
Страница 53: ...Routine Maintenance ...
Страница 68: ...68 Frontier IR Single range Spectrometers User s Guide ...
Страница 69: ...Advanced Maintenance ...
Страница 88: ...88 Frontier IR Single range Spectrometers User s Guide ...
Страница 89: ...Appendices ...
Страница 100: ...100 Frontier IR Single range Spectrometers User s Guide ...