200
Maintenance
Nebulizer
Check that the nebulizer is not clogged and that the sample capillary tubing is clean
and in good condition. Flush the nebulizer daily with deionized water.
Peristaltic Pump and Drain
Inspect the pump tubing before operating and after every few hours of use. Replace
the tubing when flat spots develop (approximately every eight hours of use). Refer
to the section on the peristaltic pump for the appropriate tubing for different
solvents.
When the pump is not in use, release the pressure plate and release the tubing to
prevent flat spots from forming. If you are planning to leave the pressure plate
clamped overnight, we recommended that you use the
TubingSaver
feature.
To enable to
TubeSaver
feature start up the ICP software. Go to the
Instrument
tab
->
Devices ->
Pump
from the dropdown menu. In the Pump screen select a
checkmark in the
Enable TubingSaver
box.
Always remove the sample capillary or the autosampler probe from the solution it
is in when you are finished using the pump. Otherwise, the solution will siphon into
the spray chamber and flood it.
Caution
Attention
When the humidity in your lab is high, water can condense on the plasma
induction plates when the torch is not in operation.
To prevent damage
to the plasma induction plates, use a soft dry cloth to gently dry the
plasma induction plates before igniting the torch.
Be careful not to
change the position of the plasma induction plates.
Lorsque le taux d'humidité dans votre laboratoire est élevé, l'eau peut se
condenser sur les plaques à induction de plasma lorsque la torche ne
fonctionne pas. Pour éviter d'endommager les plaques à induction de
plasma, utilisez un chiffon doux et sec pour sécher doucement les plaques
à induction de plasma avant d'allumer la torche. Veillez à ne pas modifier
la position des plaques à induction de plasma.
Содержание AVIO 500
Страница 1: ...AVIO 500 Customer Hardware and Service Guide ICP OPTICAL EMISSION ...
Страница 2: ......
Страница 3: ...Avio 500 Customer Hardware and Service Guide ...
Страница 12: ...Contents 10 ...
Страница 30: ...28 ...
Страница 31: ...C h a p t e r 1 ...
Страница 32: ......
Страница 33: ...Safety Practices 1 ...
Страница 58: ...56 Safety Practices ...
Страница 59: ...C h a p t e r 2 ...
Страница 60: ......
Страница 76: ...74 Preparing Your Laboratory ...
Страница 77: ...C h a p t e r 3 ...
Страница 78: ......
Страница 79: ...System Description 3 ...
Страница 94: ...92 System Description Figure 3 7 GemCone nebulizer and end cap N0680504 also shown in cross sectional view ...
Страница 115: ...C h a p t e r 4 ...
Страница 116: ......
Страница 137: ...135 Installing the Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 152: ...150 Installation Figure 4 22 Scott Spray Chamber Clip Assembly in Place ...
Страница 195: ...193 Shipping Lists 09290798 USB Cable Table 4 6 Spares Available Continued ...
Страница 196: ...194 Installation ...
Страница 197: ...C h a p t e r 5 ...
Страница 198: ......
Страница 209: ...207 Quick Change Adjustable Torch Module Figure 5 1 Quartz Torch Item Description 1 Plasma Induction Plates 2 Torch 2 1 ...
Страница 217: ...215 Quick Change Adjustable Torch Module 1 2 10 8 9 3 4 7 6 5 ...
Страница 316: ...314 Maintenance ...
Страница 317: ...C h a p t e r 6 ...
Страница 318: ......
Страница 339: ...337 RF Generator Sample Introduction System Troubleshooting Form Figure 6 2 Normal plasma at 3mm position ...
Страница 340: ...338 Troubleshooting Figure 6 3 Rounded Plasma due to sample or air getting around the outside typically subtle air leak ...
Страница 342: ...340 Troubleshooting Figure 6 5 Injector too far forward ...
Страница 343: ...341 RF Generator Sample Introduction System Troubleshooting Form Figure 6 6 No Aux Gas torch may be glowing ...
Страница 344: ...342 Troubleshooting Figure 6 7 Air leak or spray chamber temperature too high ...
Страница 345: ...343 RF Generator Sample Introduction System Troubleshooting Form Figure 6 8 Thin plasma due to leak in plasma gas line ...
Страница 346: ...344 Troubleshooting ...
Страница 354: ......