XRpad2 4336
14
68153 01
D
IGIT
A
L
I
MAGING
includes the specification for the XRpad2 LBP.
8.5.1.1
Charging the XRpad2 LBP
•
A new rechargeable Lithium Battery Pack (XRpad2 LBP) comes in a discharged condition
and must be charged using the dedicated lithium battery charger (XRpad LBC) before use.
Refer to the XRpad LBC manual for more details.
•
The XRpad LBC will charge the XRpad2 LBP to a usable condition within three hours
depending on the initial state of charge. The X-ray a detector, when connected to the XRpad
Interface and Power Unit (XRpad IPU-2), can also charge the XRpad2 LBP, but the charge
rate is much slower.
•
A charged battery will eventually lose its charge if unused. Upon initial use (or after a
prolonged storage period), the battery may require three to four charge/discharge cycles
before achieving maximum capacity.
•
The actual battery run-time will depend on the power demands made by the X-ray detector.
•
The XRpad2 LBP is keyed and can only be inserted into the XRpad LBC charger in one
orientation.
Table 9
Specification of the XRpad2 LBP
General
REF
95510949H-01
Voltage
11.1V
Amp-hours
4.2Ah
Capacity
46.6Wh
Charging Time
Within 3 hours
Environmental
Relative Humidity
35% to ~85%
Operating (discharging)
-10°C to 55°C
Charging
0°C to 42°C
Transportation
-20°C to 45°C
Storage (up to 3 months)
15°C to 35°C
Ingress
Ingress Protection Rating
IP54
Lifetime
Charge-discharge Cycles
500 cycles on average under normal usage conditions.
Battery should be discarded on or before five years from
date of manufacture indicated on the battery label.
Содержание XRpad2 4336
Страница 14: ...This page intentionally left blank xiv www perkinelmer com 68153 01 List of Figures XRpad2 4336 ...
Страница 16: ...xvi www perkinelmer com 68153 01 List of Tables XRpad2 4336 This page intentionally left blank ...
Страница 62: ...This page intentionally left blank XRpad2 4336 46 68153 01 ...
Страница 66: ...This page intentionally left blank 50 www perkinelmer com 68153 01 Index XRpad2 4336 ...