![Perkin-Elmer HATR Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/perkin-elmer/hatr/hatr_sampling-accessory-users-manual_1530154011.webp)
HATR Sampling Accessory User’s Guide . 11
Figure 7 HATR installed in a Frontier IR System
NOTE: When the HATR is installed, a connection is made between the accessory and the
purge system used in the instrument. To purge the accessory, use the left-hand
(Sample) purge connector on the rear of the instrument. The right-hand purge
connector purges the instrument.
Accessory Detection by Spectrum Software
NOTE: If you do not have an instrument installed, see the Administrator’s Guide for your
software for details of the Instrument Install Wizard.
The Spectrum software detects the presence of the accessory, and the HATR icon
is
added to the toolbar. The Setup Instrument Basic (Figure 8) and Setup Instrument BeamPath
(Figure 9) tabs are updated to show that the HATR Accessory is in position.
Figure 8 Setup Instrument Basic tab with HATR