Страница 1: ...RTD THERMOMETER TC0370 Instruction Manual...

Страница 2: ...3 Selecting the Temperature Scale 4 4 Data Hold Operation 4 5 Back light Operation 4 6 Relative Operation 4 7 MAX MIN AVG Operation 4 8 Auto Power Off 4 9 Low Battery Condition 5 Temperature Measureme...

Страница 3: ...1 INTRODUCTION P 2 This instrument is a digital thermometer for use with platinum type temperature sensor Temperature indication follows IEC751 temperature table for PT type sensor...

Страница 4: ...EASUREMENT CURRENT Approx 0 53mA MAXIMUM VOLTAGE AT TEMPERATURE COUPLE INPUT 60V DC or 24Vrms AC ENVIRONMENTAL Operating Temperature and Humidity 0 C 50 C 32 F 122 F 0 80 RH Storage Temperature and Hu...

Страница 5: ...ng 0 06 F Note The basic accuracy Specification does not include the error of the probe please refer to the probe accuracy specification for additional details SAMPLE RATE 2 Times Per Second WATER RES...

Страница 6: ...Type Indication This indicates that the display data is being hold The Maximum value is now being displayed The Minimum value is now being displayed The Average value is now being displayed The readin...

Страница 7: ...BUTTON LOCATION P 6 1 2 3 4 5 6 7 8 Pt type temperature sensor connector LCD display Power ON OFF button HOLD button 2 3 4 1 MAX MIN Average control button Back light button Relative readout button C...

Страница 8: ...by pressing C F button and vice versa to Celsius 4 4 Data Hold Operation The user may hold the present reading and keep it on the display by pressing the HOLD button When the hold data is no longer n...

Страница 9: ...of latest 8 data is kept in the memory simultaneously and updated with every new data When the MAX symbol is displayed the Maximum is shown on the display Press again then the MIN symbol is on the di...

Страница 10: ...P 9 4 8 Auto Power Off By default when the meter is powered on it is under auto power off mode The meter will power itself off after 30 minutes if no key operation One may press and hold HOLD button...

Страница 11: ...rature sensor is located at the end of the metal sheath of the sheath type TEMPERATURE PROBE To accurately test internal temperature insert the probe into the item you want to measure to a distance at...

Страница 12: ...attempting following SETUP Setup mode is cancelled during below procedure if POWER button is pressed P 11 6 TEMPERATURE METER CALIBRATION SETUP Below table diagram shows each button function when use...

Страница 13: ...uttons for 2 seconds to enter setup mode 6 1 2 Press Hold button to enter PT selection mode 6 1 3 Press Max button to select between PT probe type Pt100 Pt500 Pt1000 is circulate 6 1 4 Press C F butto...

Страница 14: ...mode 6 2 2 Press REL button to enter calibration mode Enter PT selection first 6 2 3 Press C F button to confirm PT selection 6 2 4 Press REL button to enter calibration mode Insert temperature probe...

Страница 15: ...confirm present calibration value 6 2 6 Press C F button to confirm selection 6 3 Recall Default factory setting value 6 3 1 Press and hold Power Light buttons for 2 seconds to enter setup mode 6 3 2...

Страница 16: ...BRATION SETUP 6 3 3 Press C F button to confirm PT selection 6 3 4 Press Light button to enter recall mode 6 3 5 Press C F button to confirm revert back to Default factory setting value 6 3 6 Press C...

Страница 17: ...tteries 7 1 2 Fit cover properly and tighten screw The unit s back cover are fitted with rubber rings After replacing the battery check that the rubber rings are properly seated before reinstalling th...

Страница 18: ...with a clean lint free antistatic and dry cleaning cloth Do no use cleaning agents that contain carbon or benzenes alcohol or anything similar to clean the product since these substances damage the su...

Страница 19: ...Platinum resistance thermometer sensor Pt 100 4 wires IEC751 class A 0 15 C 0 002t t measurement temperature 100 to 400 C 148 to 752 F Approx 3 2mm 0 125 Approx 120 mm 4 72 Approx 1100 mm 43 3 EN6052...

Страница 20: ...r products and services please send us an email cs perfectprime com For B2B or project based application please send an email sales perfectprime com FOR MORE INFORMATION ABOUT PERFECTPRIME PLEASE VISI...

Страница 21: ...Tyche Smart Limited cs perfectprime com 2nd Floor 107 Charterhouse Street EC1M 6HW London England UNITED KINGDOM 44 203 7695377 Retailer Email Address Telephone...