(2) FRONT PANEL
No.
NAME
FUNCTION
1
Lamp Switch
ignites the main lamp. The LED lights green when switched ON. When the main lamp
fails to ignite, the LED flashes. Press the lamp switch again to ignite the auxiliary lamp.
-Aside from the pre-use inspection of the equipment, the lamp in the video
processor should be turned OFF when the video system is not clinically used.
-The auxiliary lamp is incorporated to the processor with serial number beginning
UB and EB.
2
Lamp Life Indicator
indicates hours for Xenon lamp installed in video processor. If the indicator lights red, a lamp
should be replaced before beginning the next procedure.
3
AUTO/XLUM Select Switch
selects AUTO (automatic) or XLUM (manual) brightness control mode. AUTO or XLUM
indicators will light to indicate which is selected.
4
AVE/PEAK Select Switch
Selecting AUTO will require selection of light measuring method, AVERAGE or PEAK.
AVE or PEAK indicator lights to indicate which is selected.
AVERAGE: the brightness level is adjusted with respect to an averaging of the brightness
of the video signal.
PEAK:
the brightness level is adjusted with respect to the brightness of the peak of
the screen.
5
Brightness Indicator
indicates the brightness level settled by the user.
6
Brightness
controls the brightness level.
Up or Down button will change the brightness level as shown on the brightness indicator.
Adjustment
△
increases the brightness level.
Switch
▽
descreases the brightness level.
-Minimum reguired brightness should be used at all times to avoid risk of injury
to the patient.
7
Color Balance Switch
adjusts the video image color, Blue or Red by +/- 5 steps.
8
Pump on/off
controls air pumps on/off. LED on the switch will light when switched ON.
9
Pump high/low
LED on the switch indicates the pump output pressure level, High/Low.
WARNING
-When selecting HIGH pressure, be careful not to deliver too much air.
10
White Balance Switch
adjusts the white balance of the video endoscope. After adjustment, “WB OK!” is
displayed for about 3 seconds.