PENTAX VIDEO DUODENO SCOPE

ED-3490TK

OPERATION MANUAL

This manual describes the recommended procedures for inspecting and preparing 
the equipment prior to its use. For cleaning, high-level disinfection, and sterilization, 
refer to the separate Instructions for Use (Reprocessing) with the model name of the 
instrument.

Only for the US

Содержание ED-3490TK

Страница 1: ...l describes the recommended procedures for inspecting and preparing the equipment prior to its use For cleaning high level disinfection and sterilization refer to the separate Instructions for Use Rep...

Страница 2: ...que or any medical aspects regarding the use of the equipment For the cleaning and maintenance after its use please refer to the separate Instructions for Use Reprocessing The text contained in this m...

Страница 3: ...USE 5 2 1 INSPECTION OF THE VIDEO PROCESSOR 5 2 2 INSPECTION OF ENDOSCOPE 6 2 3 PREPARATION JUST BEFORE INSERTION OF ENDOSCOPE 15 3 DIRECTIONS FOR USE 16 3 1 PRETREATMENT 16 3 2 INSERTION AND WITHDRAW...

Страница 4: ...STRAIN RELIEF RUBBER INLET SEAL OF B190 Allows passage of accessories while preventing escape of fluids and air CONTROL BODY INSTRUMENT CHANNEL INLET For introduction of biopsy forceps and other acces...

Страница 5: ...on Align the black arrow on the soaking cap with the green dot at the base of the silver collar surrounding the electrical contacts on the PENTAX PVE connector Press the cap down onto the metal collar...

Страница 6: ...ter of an endoscopic accessory instrument must be at least 0 2 mm less than the specified instrument channel diameter in PENTAX endoscopes Working length of an endoscopic accessory instrument may be a...

Страница 7: ...uency of use In which case the lamp life might become shorter than its respective rated hours EPK i 500 hours EPK 1000 400 hours CAUTION Please refer to the instruction supplied with the processor NOT...

Страница 8: ...PENTAX video processor model EPK i or EPK 1000 CAUTION To avoid discontinuation of the endoscopic procedure have an extra spare instrument available as a standby device Should any unforseen event or...

Страница 9: ...h may be left idle and uncovered for a prolonged period of time should not be used during any inspection testing of the endoscope CAUTION Various types of endoscope leakage testers exist including man...

Страница 10: ...ials unique parts and intricate components with strict dimensional tolerances Specialized assembly techniques and application of specific sealants and or adhesives are required to ensure the watertigh...

Страница 11: ...on of the deflection controls may be an early indication of internal damage to and or part s failure within the endoscope s angulation system To avoid the possibility of further endoscope damage or th...

Страница 12: ...the end user replace the entire A W valve with a new one if the check valve is damaged missing WARNING Worn or damaged valves O rings in particular should be replaced with new ones which have already...

Страница 13: ...large drops and or squirts of oil directly onto the metal valve stems instead simply place a small droplet of oil on one s sterile gloved forefinger and gently swirl between thumb and forefinger Next...

Страница 14: ...ected For easier identification a orange colored indicator is placed on top of the OF B120 valve mechanism Remove the valve from the control body and make sure that rubber portions of the valve are no...

Страница 15: ...not move smoothly or feels sticky remove the valve from the suction cylinder on the control body of the endoscope Apply a small amount of silicone oil lubricant OF Z11 onto rubber part and the rubber...

Страница 16: ...nding radius Please note that damage to the endoscope and or accessory instrument is possible if excessive force is applied during insertion or withdrawal of the EAI To prevent equipment damage or dev...

Страница 17: ...please read carefully and follow all instructions in the manual supplied with the fluids for use and pay special attention to any reactions with the materials identified in the intended fluid path NOT...

Страница 18: ...and disinfected or sterilized after any previous use and after being returned for any repairs service Refer to the companion Instructions for Use Reprocessing describing in detail PENTAX reprocessing...

Страница 19: ...use thermal injury To minimize the risk use only the minimum amount of brightness and avoid close stationary viewing and unnecessary prolonged use CAUTION ANY lack of smooth operation of the deflectio...

Страница 20: ...flation may exist from one manufacturer s equipment light source video processor and or scope type to another It is therefore important to closely monitor the patient at all times and to aspirate exce...

Страница 21: ...ssive pressure when introducing any accessory since the instrument channel may be damaged Malfunction of the scope as well as costly repairs may result 4 When a portion of the cups of the forceps beco...

Страница 22: ...repeated short strokes to advance the cannula 2 Attach a luer lock syringe filled with contrast material to the cannula Inject until air is eliminated from cannula This will maintain the integrity of...

Страница 23: ...nd the connecting socket of the patient ground of the electro surgical generator In case of Floating Ground Electro Surgical Generator without a scope feedback cord Do not use optional functional grou...

Страница 24: ...L Z3 intended to reduce potential RF interference and electronic noise that may appear in the endoscope image when using electro surgical devices Ensure that cable OL Z3 is correctly connected between...

Страница 25: ...unauthorized PENTAX entities which still bear a PENTAX label are within PENTAX device specifications and or if unauthorized activities have significantly changed the instrument s performance intended...

Страница 26: ...n 120 90 Right Left 105 90 Rigid Distal Width 13 2mm Distal End Width 13 2mm Insertion Tube Width 11 6mm Maximum Insertion Portion Width 13 4mm Minimum Instrument Channel Width 4 2mm Insertion Tube Wo...