DESCRIPTION
12
3.9.4
Ventilator Power Supply
The mains lead for an AV-series ventilator
can be plugged into one of the auxiliary
power sockets on the rear of the machine.
AV-S with interface link to Prima SP2
a)
Turn the machine Gas Delivery Switch ON.
The ventilator will power-up.
b)
While the Prima SP power is ON, the
Ventilator can be turned OFF and ON, using
the ventilator On/Off switch.
c)
Turn the Gas Delivery Switch to OFF.
The ventilator will power-down.
For further information, refer to section 3.6 in the
AV-S user manual.
AV-S Ventilator Back-up Battery
If the power supply to the ventilator fails, the
ventilator back-up battery will power the
ventilator for 30 minutes, if the battery has
been maintained in a fully charged condition.
Refer also to the user instruction manual
supplied with the ventilator
Battery charging takes place automatically
when the ventilator mains lead is connected
to a ‘live’ mains supply.
The OFF indicator on the ventilator front
control panel will show a yellow light during
charging.
NOTE
The stated battery back-up period will only be
available if the battery is kept fully charged.
If the battery has been allowed to discharge
below the LOW BATTERY condition, the
ventilator will not function correctly until the
voltage raises above the LOW BATTERY level.
A fourteen hours recharge will be necessary to
bring the battery to full charge.
3.9.5
Monitor and other Accessories
The mains lead (or adaptor) for a monitor
system or other accessories requiring an
electrical supply can be plugged into one of
the auxiliary sockets on the rear of the
machine.