Latest training information can be found on: 

https://portal.e-lfh.org.uk/Component/Details/606835

24-hour helpline number: 

01455 849199

This quick start guide has been developed by the NHS and Cabinet Office for the Ventilator Challenge programme and should be use

d in conjunction with the manufacturer’s Instructions for Use 

RESTRICTED DEVICE ONLY for use in COVID-19 pandemic
ONLY to be used for emergency ventilation.

• ALL adverse incidents must be reported to MHRA on 0800 731 6789 
• Manual back up ventilation must be available 

Follow instructions for use

1

2 3

4

5

6

7

8

Display screen

Oxygen 

set to 9

Air 

set to 3

Bacterial filter

HMEF filter

On switch

Bellows 

Dial (turn to select, push to confirm)

This quick start guide has been developed by the NHS and Cabinet Office for the Ventilator Challenge programme and should be use

d in conjunction with the manufacturer’s Instructions for Use 

This machine does not use an absorber

A

Pressure Controlled 
Ventilation 

(PCV) (image 

A

)

4

How do I change settings

5

How do I check 
the device?

• A  P.O.S.T. will run automatically 

when the device is switched on

• To run a function test, select 

ADULT DEFAULTS setting, 
connect a test lung, fill bellows 
with oxygen, select VOLUME 
CYCLE mode

• After 10 breaths check the 

bellows returns to the top of the 
housing. Check on the  screen 
that the tidal volume is between 
600ml- 800 ml

• Remove the test lung and 

occlude the patient ‘Y’ piece to 
check the HIGH AIRWAY 
PRESSURE alarm; unocclude
the patient ‘Y’ piece to check the 
LOW AIRWAY PRESSURE 
alarm

6

D

This machine does not use an absorber

OPERATE THE DEVICE

PENLON ESO2: 

QUICK START GUIDE

TYPE B ANAESTHETIC MACHINE TYPE

V

1.4

5

0

0

9

4

5

7

 

What do the 
alarms mean?

1. Pressure jumps excessively 

at beginning of inspiration:

• Airway obstruction OR  Flow rate is 

too high; decrease TV setting

2. Pressure jumps excessively 

at end of inspiration:

• TV setting is too high; decrease TV 

frequency setting OR,

• Could be caused by mucous 

plugging; patient requires 

suctioning.

*

3. Sudden increase in patient 

airway pressure:

• Tube kinking, coughing, 

asynchrony or other obstructions

4. Inflation pressure too low:

• Leak in the circuit or disconnection; 

check hose or connections 

• Check ventilation parameters and 

that patient valve is functioning

7

*

Procedures such as suctioning will cause a loss of pressure 

within the circuit and emptying of the bellows 

• To offset, increase the oxygen flow by turning the flow  knob (IMAGE 

A

anticlockwise past the point that the bobbin is at the top of the flow tube 

• Return to 9L/min immediately after reconnection and check ventilation

• Use the touch screen 

or the dial 

pressing to confirm 

• In the 

ADULT DEFAULTS 

setting, 

VOLUME

mode, the parameters that can be changed 
are: 

frequency, tidal volume, pressure limit, 

I:E ratio, PEEP

• Use the display screen to monitor parameters. 

The default waveform is always Pressure vs. 
Time (cmH2O vs. seconds) (image 

D

)

• Start in 

Volume Controlled Ventilation 

(Section 3)

• Select ‘

PRESS’ 

on the display screen 

and press once

Set TARGET to 35 cmH

2

• Select by pushing grey knob 

Ventilation starts

• Set 

Rate to 20, I:E 1:2, PEEP to 15 

using 

(with a compliant test lung / bag 

large tidal volumes will be delivered)

• Check Oxygen reads 

80%

• Attach to patient AND Check Tidal Volumes

Procedures requiring circuit disconnection

8