This quick start guide has been developed by the NHS and Cabinet Office for the Ventilator Challenge programme and should be use

d in conjunction with the manufacturer’s Instructions for Use 

This quick start guide has been developed by the NHS and Cabinet Office for the Ventilator Challenge programme and should be use

d in conjunction with the manufacturer’s Instructions for Use 

How do I connect the ventilator? 

(Image 

A

)

Connect medical gas hoses to oxygen 

and air and connect to mains power

Connect breathing system to 

and 

Switch 

on, indicator turns green

1

How do I set up the breathing 
system and the HME filter? 

(Image 

B

)

• Attach catheter mount to the HME filter 

and then to the Y piece of the breathing 
circuit

• Ensure that the other end of the system is 

attached to 

A

and 

A

with filters as 

shown 

B

2

SET THE CONTROLS

Catheter Mount

HME Filter

Volume Controlled Ventilation

(Image 

A

C

) START UP MODE

• Turn device 

On

• Set 

Oxygen to 9L/min 

• Set 

Air to 3L/min

• Attach patient end of breathing system to 

test lung / bag observe  bellows fill

• Select ‘ADULT DEFAULTS’ on the display 

screen 

• Select ‘VOLUME’ on the display screen 

and press once

Set Volume to 400 mls, Rate to 20, 
I:E = 1:2 and PEEP to 15

• Press 

Volume

twice to start ventilation

• Check Oxygen reads 

80% and

High 

pressure alarm at 

35 cmH

2

O

• Only then attach to patient

3

B

ADULT DEFAULT screen view

C

1

2 3

4

5

6

7

8

Display screen

Oxygen 

set to 9

Air 

set to 3

Filter

Filter

On switch

Bellows

Dial (turn to select, push to confirm)

A

PENLON ESO2: 

QUICK START GUIDE

This machine does not use an absorber

CONNECT AND SET UP THE DEVICE

TYPE B ANAESTHETIC MACHINE TYPE