This quick start guide has been developed by the NHS and Cabinet Office for the Ventilator Challenge programme and should be use
d in conjunction with the manufacturer’s Instructions for Use
This quick start guide has been developed by the NHS and Cabinet Office for the Ventilator Challenge programme and should be use
d in conjunction with the manufacturer’s Instructions for Use
How do I connect the ventilator?
(Image
A
)
•
Connect medical gas hoses to oxygen
and air and connect to mains power
•
Connect breathing system to
and
•
Switch
on, indicator turns green
1
How do I set up the breathing
system and the HME filter?
(Image
B
)
• Attach catheter mount to the HME filter
and then to the Y piece of the breathing
circuit
• Ensure that the other end of the system is
attached to
A
and
A
with filters as
shown
B
2
SET THE CONTROLS
Catheter Mount
HME Filter
Volume Controlled Ventilation
(Image
A
&
C
) START UP MODE
• Turn device
On
• Set
Oxygen to 9L/min
• Set
Air to 3L/min
• Attach patient end of breathing system to
test lung / bag observe bellows fill
• Select ‘ADULT DEFAULTS’ on the display
screen
• Select ‘VOLUME’ on the display screen
and press once
•
Set Volume to 400 mls, Rate to 20,
I:E = 1:2 and PEEP to 15
• Press
Volume
twice to start ventilation
• Check Oxygen reads
80% and
High
pressure alarm at
35 cmH
2
O
• Only then attach to patient
3
B
ADULT DEFAULT screen view
C
1
2 3
4
5
6
7
8
Display screen
Oxygen
set to 9
Air
set to 3
Filter
Filter
On switch
Bellows
Dial (turn to select, push to confirm)
A
PENLON ESO2:
QUICK START GUIDE
This machine does not use an absorber
CONNECT AND SET UP THE DEVICE
TYPE B ANAESTHETIC MACHINE TYPE