4.5 Functional
Tidal Volume
Adult bellows
20 to 1600 ml (±10%)
Paediatric bellows
20 to 350 ml (±10%)
At ambient temperature of 20oC (+/-10%) and ambient atmosphere of 101.3 kPa (+/-10%)
.
Minute Volume 0 to 30 L
Rate
4 - 100 bpm
I:E Ratio
1:0.3 - 1:8
Pressure Limit
10 - 80 cmH
2
O
Fresh Gas Compensation
Automatic Tidal Volume adjustment
Modes
Off
Standby
Volume Cycle
Pressure Controlled
Spontaneous (includes advanced breathing modes)
Pressure Control
10 - 70 cmH
2
O
Inspiratory Flow
2 - 75 L/min
Spontaneous Mode
Active Volume and Pressure Alarms,
Advanced Breathing Modes selectable
(see section 4.6)
Electronic PEEP
4 - 30 cmH
2
O
Oxygen Monitor
Fuel Cell type
For full specification, see section 4.15.
Spirometry - Resolution
±
1 ml
Ventilator Performance - accuracy of delivered volumes
>300 ml
± 10%
>100 ml <300 ml
± 20%
<100 ml
± 50%.
NOTE
The ventilator is designed for use with Spirometry ON.
Accuracy of delivered volumes with Spirometry OFF may vary from the figures given above.
SPECIFICATION
35
Содержание AV-S