PendoTECH PressureChecker PDKT-650-905 Скачать руководство пользователя страница 8

PendoTECH PressureChecker User Guide 

 

 

Revision 1 

 

 

 

Page 8/15

 

 

2. Using the Device  

 

2.1.

 

Testing sensors non-invasively 

Positive and vacuum pressures can be generated within PressureChecker and applied to a sensor 
under test. The pressure delivered is indicated on PressureChecker’s liquid crystal display for quick 
comparison to the monitor. By applying a vacuum to the sensor test port, the linearity and accuracy of 
the pressure sensor can be tested. Since the pressure sensor is a true differential device (measures 
gauge pressure relative to atmosphere), applying vacuum is equivalent to applying a pressure to the 
fluid path so the system can be checked to verify calibration very easily without accessing the fluid 
path or breaking sterility if a sterile fluid path. Even though it may not be possible to test the entire 
range of the pressure sensor, the proper function and sensor output accuracy of 0.2584 mV/psi/V can 
be verified to be within accuracy specification because this applies across the range of sensor 
pressures to 30 psi.  Beyond the 30 psi, there is a non-linear output that both the PressureMAT and 
PendoTECH Control Systems can interpret and display the actual pressure reading.  Regardless, the 
PressureChecker only can be used in its specified range which is below 30 psi. 
 

NOTE:  During the following steps the fluid path of the sensor must be exposed to atmosphere 
 

Steps:  
1.

 

Connect the sensor to the monitor cable 

2.

 

Connect the 4 foot air tube with stop cock and slip luer on one end to the sensor test port and the 
PressureChecker pneumatic port on the other end. 

3.

 

Zero the system- properly zeroing the system is important for accuracy determination.  To zero do 
the following: 

a.

 

Position the handle on the stopcock to vent both the sensor and pneumatic port to 
atmosphere. 
 

 

 

b.

 

Tare the sensor at the monitor per its operating instructions 

c.

 

If the PressureChecker display does not read 0.00, turn the zero adjustment located on the 
right side using a small screwdriver 
 
 
 

Содержание PressureChecker PDKT-650-905

Страница 1: ...PendoTECH PressureChecker Model PDKT 650 950 Revision 1 www pendotech com User Guide...

Страница 2: ...or what is covered in the Limited Warranty and Terms and Condition of Sale WARNING WARNING is used to indicate the presence of a hazard which can cause severe personal injury death or substantial prop...

Страница 3: ...6 1 2 3 Theory of Operation 7 2 Using the Device 8 2 1 Testing sensors non invasively 8 2 2 Using as a monitor simulator to test a monitor 11 3 Key System Components 13 3 1 Power Supply and Low Batter...

Страница 4: ...the performance of the sensor by the end user PendoTECH Pressure Sensors are also frequently used on single use bioreactors to measure pressure in the bag The required resolution is often less than 1...

Страница 5: ...atic port provides the external connection for fittings to connect to the sensor test port Verify proper functioning of sensor and monitors Verify proper function of sensor in situ by accessing the at...

Страница 6: ...PendoTECH PressureChecker User Guide Revision 1 Page 6 15 The control functions are on the right side panel as shown below Adjustment Knob Zero Adjustment ON OFF Switch 1 2 2 Specifications...

Страница 7: ...pressures The pressure sensor simulator which is housed inside PressureChecker s monitor connector is a pressure sensor of the same type as that used in the PendoTECH Single User Pressure Sensor When...

Страница 8: ...uracy of 0 2584 mV psi V can be verified to be within accuracy specification because this applies across the range of sensor pressures to 30 psi Beyond the 30 psi there is a non linear output that bot...

Страница 9: ...STING DO NOT EXCEED 7 PSI OR THE SENSOR UNDER TEST MAY BE DAMAGED WARNING DURING THIS TESTING DO NOT APPLY LESS THAN 7 PSI OR THE PRESSURE SENSOR SIMULATOR MAY BE DAMAGED 4 As pressure is generated by...

Страница 10: ...ine the maximum allowable deviation in pressure d Compare the results from step b with step c If the results from b fall into the range calculated in c the sensor under test is within the specificatio...

Страница 11: ...ith stop cock from the PressureChecker pneumatic port and place the cap on the pneumatic port and tighten in place or the stop cock may be closed by turning handle parallel with tubing and positioned...

Страница 12: ...ation on each item in the setup monitor PressureChecker and pressure sensor simulator determine the maximum allowable deviation in pressure d Compare the results from step b with step c If the results...

Страница 13: ...sure is from 7 to 10 5 psi 7 psi if a sensor test port is connected to pneumatic port because the creates a vacuum on the sensor under test and 7 psi is the sensor lower limit The maximum allowable ov...

Страница 14: ...ite travel and diagnosis under this warranty clause shall be borne by Buyer unless accepted in writing by Seller Failure by Buyer to give such written notice of defects within the applicable time peri...

Страница 15: ...should be returned for repair The PressureChecker should be checked for calibration every twelve months or as often as company policy requires The calibration service is available from PendoTECH for...

Отзывы: