Pelton & Crane Solaris Скачать руководство пользователя страница 1

Solaris

Sterilization Center
Use & Care Manual

Содержание Solaris

Страница 1: ...Solaris Sterilization Center Use Care Manual ...

Страница 2: ...ration stage 1 continued 10 operation stage 2 11 operation stage 3 12 operation stage 4 13 CLEANING DISINFECTING STERILIZATION 14 Vacuum Instruments 15 Cleaning Cabinetry Surfaces 16 ADJUSTMENTS 17 Notes 18 Technical Support Technical assistance is available Monday through Friday 8 00 am to 6 00 pm Eastern Standard Time Phone 800 659 5922 Fax 800 659 7255 Customer Service 800 659 6560 ...

Страница 3: ...AGE TOWER STAGE 4 STERILE STORAGE STAGE 3 STERILIZATION STAGE 2 PREPARATION AND PACKAGING STAGE 1 RECEIVING CLEANING DECONTAMINATION S T A G I N G A R E A S The Solaris Sterilization Center is uniquely designed into four staging sections that allow the user to keep each process of sterilization separated Each stage station is divided by a partition to reduce cross contamination and to improve man ...

Страница 4: ...be used It is the user s risk when using non released accessories are used Obtaining Technical Literature The manufacturer will make available on request circuit diagrams component parts lists descriptions calibration instructions or other information that will assist technical personnel to repair and replace serviceable items Product Identification This dental cabinet can be identified by its pro...

Страница 5: ...connected only to supply mains with protective earth See Grounding Instructions WARNING This product is intended for use by dental medical professionals only WARNING Only authorized service technicians should attempt to service this equipment Serviced by other than authorized technicians will void the warranty WARNING Power cords and their associated parts cannot be substituted without increased r...

Страница 6: ... to legal regulations alterations to the cabinet which could impair the safety of the operator patient or third parties are unlawful For reasons of product safety only original Pelton Crane accessories approved for this product or accessories by third parties which have been approved by Pelton Crane may be used Using non approved accessories is at the risk of the user general CAbinet information T...

Страница 7: ...e 3 This outlet must be hard wired to accommodate 120V or 230V outlets 3 The fourteen LED lights 6 White 4 Blue and 4 Red located in Stages 1 2 and 4 are automatically activated when the control panel s power swtich is ON The lights will remain on until the control panel power is de activated WARNING The dental sterilization center system is designed for use in a dental environment and must be ope...

Страница 8: ... and have integral shut off valves 80 PSI is the maximuum air pressure Do not exceed the maximum air pressure 5 Optional pneumatic doors in Stage 1 can be opened by pressing in the foot pedal at the base of the cabinet The doors will close when the pedal is pressed a sec ond time Operation MODE SOL ARIS STAGE 1 STAGE 2 STAGE 3 STAGE 4 4 4 5 5 ...

Страница 9: ... located in the base to store frequently used supplies Ultrasonic station may be recessed in the base cabinet ultrasonic not included False panel can be utilized for accessories 4 The 14 X 18 X 10 stainless steel sink accommodates large bulk cassettes that can be difficult to clean An HVE cannister and an integrated drain board are equipped in the sink module to assist in clean up 5 The sink modul...

Страница 10: ... flow 10 The water flow is controlled by the faucet s handle Turning the handle right activates the cold water and turning it to the left activates the hot water Pulling the handle back increases water pressure Pulling the handle forward decreases pressure 11 The VistaPure wand provides high purity water for the water bottles autoclave and office drinking water It is attached to the filtration sys...

Страница 11: ...d bags for individual instrumentation 3 The drop down allows for easy access to supplies such as tape indicator strips and a heat and seal An electrical outlet is incorporated into this module for optional equipment that requires an electrical resource 4 An upper storage and lower storage modules provide extra space for storing essentials such as cassettes and packaging materials that are needed w...

Страница 12: ...lizer directly into the reservoir drain eliminating the use of a bucket 3 The storage area in the base of the cabinet provides an area to store distilled water or a steam collector 4 An upper storage module provides extra space for storing essentials that are needed when preparing for sterilization No upper storage will be provided for the 3 shelf tower to accommodate for the space needed for two ...

Страница 13: ...ls 1 The four viewing panels allow easy inspection of contents 2 Interior lighting for easy identification of sterile cassettes 3 Five adjustable racks to accommodate all varieties and sizes of cassettes operation stage 4 1 2 3 WARNING Use caution when pulling the sterile storage tower out Pulling the tower out too quickly could result in the tower coming off its tracks and falling over ...

Страница 14: ...endations published by the American Dental Association for further information The question is often asked What should I use to disinfect my dental unit chair and light This question is more complex than it seems because of the wide variety of products on the market as well as formulations of the products changing to meet the needs of increased asepsis Barrier Technique Chemical Disinfection Regar...

Страница 15: ...ions verifying that the air flow adjustment is regulated by the thumb lever and that the HVE and SE are being cleaned and maintained properly Disassemble the valve and lubricate the internal parts when operation becomes stiff or sticky Clean the inner surfaces and apply a small amount of O ring lubricant to the moving parts and O rings After Each Patient At the End of Each Day Cleaning the Solids ...

Страница 16: ...nces they contain many cleaning agents can affect surfaces by dissolving etching bleaching or discoloring Carefully wipe surface with a damp cloth and use an appropriate cleaning agent to lessen potential damage Cleaning Cabinetry Surfaces Cleaning Stainless Steel Fixtures Clean surface using warm water with a mild non scratching abrasive powder such as typical household cleanser Do not use brushe...

Страница 17: ...UP DOWN Screw 3 will move the door TOWARD and AWAY from the module Drawer Adjustment The drawers are adjusted at the factory However an occasional adjustment may be required over a period of usage These adjusting screws are located in each front corner of the drawer Using a 2 posi drive driver only adjust doors as follows 1 Vertical Adjustment Lightly loosen screws 2 counterclockwise adjust screw ...

Страница 18: ...18 042281 R03 Notes ...

Страница 19: ...P N 042281 Rev 03 04 21 14 Printed in USA 2014 11727 Fruehauf Drive Charlotte NC 28273 USA We reserve the right to make any alterations which may be due to any technical improvements ...