Charging the Rechargeable Battery
Use the USB cable included in the delivery to charge the rechargeable
battery of the PEHAMED
CD mon
device. The charging voltage can be
supplied either directly from a computer USB socket or via the
pluggable power supply with USB socket supplied with the device.
NOTE!
Only use the USB cable and pluggable power supply supplied
by PEHAMED that were specially designed for use with the LiPo
rechargeable batteries utilized in the
CD mon
device!
If you use other pluggable power supplies or USB cables, this could
damage the device and/or the LiPo rechargeable battery and cause
smoke or fire to develop!
If you use a different USB cable or pluggable power supply, this will
immediately void the
CD mon
guarantee!
LiPo rechargeable batteries are particularly sensitive against total
discharge and overcharging. This can immediately and permanently
damage such batteries. For this reason, the LiPo rechargeable battery
integrated in the
CD mon
device is equipped with a special electronic
protection module. This protection module prevents total discharging
or overcharging of the LiPo cells.
Always charge such rechargeable batteries only on a fireproof surface
and far away from combustible materials! Never recharge the device
on inflammable materials.
Always store a chemical fire extinguisher nearby in case of fire.
Never leave the
CD mon
device unattended during the recharging
process so that you can react quickly in case a problem occurs.
Before starting the recharging process, always first plug the pluggable
power supply into the wall socket or the “wide” USB plug into the USB
socket of the computer. Only after doing so, plug the mini USB plug into
the charging socket of the CD mon device.
Before disconnecting the device from the power supply, always first
remove the mini USB plug from the device.
During the charging process of the switched-off device, the display will
show the message “CHARGING”. In case you recharge the unit while
making measurements, the display will show the info “chrg” in the
lower right section next to the battery symbol.
As soon as the
CD mon
LiPo rechargeable battery has been
completely recharged, the device electronics will recognize this, the
recharging process will be automatically switched off and the
messages will disappear from the display.