S
PECIFICATIONS
Combination Warmer
▪
Operator’s Manual
▪
MN-47320
▪
Rev 1
▪
1/21
13
S
PECIFICA
TIONS
Specification Information
Transportation and
storage conditions
▪
Ambient temperature range of -40°F to +159°F (-40°C to +70°C).
▪
Relative humidity range of 10% to 95%, non-condensing.
▪
Atmospheric pressure range of 7.25 psi to 15.37 psi (50kPa to 106kPa).
Operating conditions
▪
The warmer must acclimate to the room temperature it will be placed in—24
hours is recommended.
▪
The recommended environmental temperature range is 60°F to 90°F
(15°C to 32°C).
▪
The recommended relative humidity is above 20%, non-condensing.
Standards for
electrical equipment
▪
Medical equipment listed by Underwriters Laboratories with respect to
electrical shock, fire, and mechanical hazards only, in accordance with UL
60601-1 and CAN/CSA C22.2 No. 601.1.
▪
Grounding reliability can only be achieved when the appliance is
connected to an equivalent receptacle marked “Hospital Grade.”
▪
IP-XO ordinary
Model
P-2140
Intended use
Top cavity—warming blankets
Bottom cavity—warming irrigation or injection fluids
Weight
Net: 374 lb (170 kg)
Ship: 494 lb (224 kg)
Storage cavity
capacity
Blanket cavity: 20–25 blankets (15.4 ft
3
)
Fluid cavity: 72 1-liter bags or 54 1-liter bottles
Temperature range
Blanket: 98°F to 180°F (37°C to 82°C)
Irrigation fluid (IRR): 98°F to 150°F (37°C to 66°C)
Injection fluid (INJ): 98°F to 104°F (37°C to 40°C)
Clearance
requirements
3” (76mm) from rear
1” (25mm) from top
1” (25mm) from sides
E471516
Protective Earth
Ground Symbol