Maintenance

Cleaning 

Mattress Cover 

Regular care should be maintained by daily wiping with a damp cloth or sponge, and periodic cleaning 

with a mild soap and water solution.

The mattress cover material is resistant to most medicinal-type stains, but 

may be damaged by solvents and dyes.  Remove any fluids which are 

spilled on the upholstery immediately.  Also, avoid the use of abrasive 

cleaners, steel wool, chlorinated bleaches, or solvents.  Do not use high 

pressure water jets, submerge, or sterilize.

Check the label on the mattress to determine the fabric type before using 

any cleaning solution.

Acceptable Cleaning Fluids  

Fabric Type

Cleaning Solution 

Trade Name

Cleaning Solution Type Cleaning Solution 

Manufacturer

For Vinyl-covered 

mattresses:

Galahad

Phenolic

Puritan

LPH

Phenolic

Steris Corp

Amphyl

Phenolic

Reckitt-Benckiser

One-Stroke

Phenolic

Steris Corp

Vesphene II

Phenolic

Steris Corp

For Urethane-covered 

mattresses:

A33

Quaternary

Ecolab

Blue Chip

Quaternary

S .C . Johnson Co .

Elimstaph

Quaternary

Walter G . Legge

Forward DC

Quaternary

S .C . Johnson Co

Lysol I . C

Quaternary

Reckitt-Benckiser

Omega

Quaternary

Ecolab

Pro-Tech RDI-36B

Quaternary

Central Solutions, Inc

Quat Sanitizer

Quaternary

Ecolab

Sanikleen

Quaternary

West Penetone

Virex 128

Quaternary

S .C . Johnson Co

Super SaniCloth

Quaternary/Isopropyl

Professional 

Disposables Inc

SaniCloth Plus

Quaternary/Isopropyl

Professional 

Disposables Inc

For Either type of 

mattress

Viro-Chek

Hypochlorite

Bob Barker Co .

Alcide LD

Chlorine Oxide

Alcide Corporation

Disinfectants

:  Phenolic disinfectants are the best choice for vinyl products, though properly diluted 

quaternaries are also acceptable.  Quaternary disinfectants are recommended for urethane products.  

Phenolics should be avoided on urethane.  Quaternary/Isopropyl disinfectants are recommended 

for urethanes, but not recommended for vinyls.  All mattresses may be cleaned with a 1:10 dilution 

of household bleaches containing 5.25% sodium hypochlorite as recommended by the Center for 

Disease Control in Atlanta, Georgia.  There is no harmful effect on the fabric.  Disinfectants applied 

at full concentration or in highly concentrated solutions will decrease the useful life of mattresses.  

Iodophor type disinfectants used on mattresses may result in staining.

Содержание P-172

Страница 1: ...P 172 P 172 Part Number 504 0060 00 Revision 08 2020 Installation Operation Manual Shown with some accessories ...

Страница 2: ... Unpacking 5 Description 6 Introduction 6 Features 6 Components Overview 6 Controls and Indicators 7 Checklist of Installed Options Authorized Accessories 8 Stretcher Operation 9 Braking and Steering Controls 9 Side Rails Option 10 Fowler Section Positioning Option 11 Locating Sockets for I V Poles Option 12 Maintenance 13 Cleaning 13 Preventive Maintenance 14 Scheduled Maintenance 15 Calling for ...

Страница 3: ... will need to provide for servicing key information is highlighted Date of Purchase Serial Number Name of Owner Facility Department Weight Capacity Name of Authorized Pedigo Dealer Telephone of Authorized Pedigo Dealer Address of Authorized Pedigo Dealer ...

Страница 4: ...center or any examination treatment or in procedures environment This could be before during and or after diagnostic therapeutic interventional medical and or surgical procedures examinations and or treatment Safety Instructions The primary concern of Pedigo is that this equipment is operated and maintained with the safety of the patient and staff in mind To assure safer and more reliable operatio...

Страница 5: ...could result in minor or moderate injury It may also be used to alert against unsafe practices Safety instructions or equivalent signs indicate specific safety related instructions or procedures Notice is used to address practices not related to physical injury This includes instructions relative to preventing equipment damage and proper function Transportation and Storage Conditions Ambient Tempe...

Страница 6: ... be kept dry Indicates the proper shipping orientation for the product 1 Carefully remove the packaging from the stretcher 2 Flip the stretcher over and allow caster wheels to rest firmly on the floor Inspect the stretcher for any shipping damage Report any damage to the shipping company and fill out a concealed damage report General items to inspect for shipping damage are 1 Stretcher does not ro...

Страница 7: ...or easier patient transfer Up to four I V rod locations for optional fixed or adjustable I V poles Blue wrap around bumpers Four corner brake steer casters Components Overview The stretcher s major components locations and names are shown below Description of Components 1 Side rails option 6 Casters 11 IV Pole option 2 Mattress option 7 Brake Pedal red 12 Shelf option 3 Litter Top 8 Steer Pedal gr...

Страница 8: ...ble below describes their function Reference Control Function 1 Fowler Handles Fowler handles allows the fowler back section to be raised or lowered See Fowler Positioning on page 11 2 Side Rail Latch Release Handles Side rail handles allow the side rails to lock in the upright position See Side Rail Option on page 10 ...

Страница 9: ...ion Number Installed Storage Shelf Stainless Steel P 170 1 Side Rails Stainless Steel P 170 2 Caster Roller Bumpers P 170 6 Option Name Factory Install NOT Required Option Number Fowler Stainless Steel P 170 3 IV Pole Fixed Height P 170 4 Restraint Straps Velcro P 170 5 Mattress Pad 2 5 thick Black P 170 8 02 Holder Requires Storage Shelf P 170 9 Bed Exit Monitor 750216 ...

Страница 10: ... or swivel preventing the stretcher from moving Disengage the steering or brakes by pressing the grey pedal on each caster Always engage steering on at least one caster when transporting patients to ensure control over the stretcher s path Also seek assistance when transporting down a steep incline to prevent loss of control of the stretcher Failure to do so could result in personal injury to pati...

Страница 11: ...e 2 away from the stretcher 3 Gently lower the side rail 1 To raise the side rail 1 Grasp the side rail 1 firmly with both hands and lift the side rail up until the latch automatically locks and secures the side rail in the up position Firmly grasp the side rail while raising or lowering Failure to do so could result in personal injury to patient or the caregiver ...

Страница 12: ...d of the stretcher 2 Lift the Fowler 1 upward slightly and move the fowler bail 2 free of the positioning fins on the litter When the Fowler 1 is in the desired position release the Fowler release handle 1 and place the fowler bail into the nearest position Firmly grasp the Fowler section with both hands during positioning Failure to do so could result in serious injury to the patient or the careg...

Страница 13: ...retcher when the I V Pole Option is installed They are located on the opposite corners one at each end of the stretcher An additional I V Pole Option can be added to make I V Pole Sockets available at all four corners The I V Pole Option includes a storage tube 2 at the head end underneath the patient surface Additional IV Pole Holder mtg locations ...

Страница 14: ... Quaternary S C Johnson Co Lysol I C Quaternary Reckitt Benckiser Omega Quaternary Ecolab Pro Tech RDI 36B Quaternary Central Solutions Inc Quat Sanitizer Quaternary Ecolab Sanikleen Quaternary West Penetone Virex 128 Quaternary S C Johnson Co Super SaniCloth Quaternary Isopropyl Professional Disposables Inc SaniCloth Plus Quaternary Isopropyl Professional Disposables Inc For Either type of mattre...

Страница 15: ...r solution for hard to clean areas and stained areas If disinfectant cleaners are used do not exceed manufacturer s recommended dilution rate or life expectancy of upholstery may be shortened Obvious damage Visually check condition of stretcher crib for obvious damage cracks in components missing components dents in components torn upholstery leaking fluids or any other visible damage which would ...

Страница 16: ...ricate latch mechanism with lithium white grease or petroleum jelly Check that latch mechanism releases and latches easily Adjust or repair latch mechanism if necessary Caster stem screws Check caster stem screws for tightness Remove any loose screws apply removable threadlock and reinstall screws and lockwashers Caster control pedal Check control pedals for proper operation Make sure all pedals w...

Страница 17: ... lbs 159 kg Overall Length 78 198 cm Overall Width 28 71 cm Height Range Fixed 33 85 cm Fowler Positioning option Fowler Backrest 3 12 26 38 44 Patient Surface 77 x 26 196 x 66 cm Side Rails 12 x 43 30 x 109 cm Casters 8 20 cm Warranty 1 year parts and labor ...

Страница 18: ...and properly cleaned as required in the Pedigo Installation and or Installation Operation Manual for this applicable product 3 products considered to be of a consumable nature 4 accessories or parts not manufactured by Pedigo 5 charges by anyone for adjustments repairs replacement parts installation or other work performed upon or in connection with such products which is not expressly authorized ...

Страница 19: ...edigo Products Inc 2013 Printed in the U S A 4000 SE Columbia Way Vancouver Washington 98661 USA www pedigo usa com TEL 800 822 3501 FAX 360 696 1700 Installation Operation Manual P 170 Series Stretchers ...

Страница 20: ...19 ...

Отзывы: