background image

Peco Services Ltd, Brough, Cumbria, CA17 4BG, England   Tel/Fax 44 (0)17683 41111 

[email protected]          www.pecoservices.co.uk 

Instructions for use 

 

Controls 

 

All of the controls are located on the front of the device above the door.  They consist of the following: 

 

 

On/Off Switch   

Located on the left hand side. Illuminates when the power is on. 

 

Fan On/Off Switch 

Located to the left of the On/Off Switch. 

 

Temperature Display 

Displays the internal air temperature of the chamber. 

 

Adjustment buttons 

Used to adjust the set-point of the digital thermostat 

 

Operation 

 

In 

Fan  Assisted

 

mode (fan switch ‘On’), the thermostat senses the air passing through the fan, and therefore a rapid warmup of the cabinet will 

occur.  The air in the cabinet is recirculated by being drawn into the fan and out of the vent in the bottom 

 

The device can be used either upright or lying down in Fan Assisted mode. 

 

In 

Convection 

mode (fan switched ‘Off’), a thermostat senses the surface temperature of the base element (with the unit lying on its back), thus 

ensuring  that  the  surface  on  which  the  animal  is  sitting  is  controlled  to  a  safe  level.    Additional  heat  is  provided  through  the  walls  of  the  device, 
however, when used in convection mode the unit will take longer to warm up than in fan assisted mode and its accuracy will be lower. 

 

!  When in Convection Mode (fan off), the device should only be used lying on its back  ! 

 

When initially switched on, the digital thermostat will perform an internal check before displaying the current internal temperature of the device.  In 

Fan Assisted

 mode the fan will also be heard to start.  Assuming that the set temperature is higher than the current internal temperature, the output 

light (‘Out’ on the display module) will illuminate, signifying that power is being supplied to the heater elements.  This light will cycle on/off with the 
thermostat. The elements are located on both sides and the floor of the internal casing, with an additional element located in the rear over which the 
air being circulated is passed.  This ensure

s that when used lying on its back, what is now the ‘floor’ of the device is not directly heated allowing 

animals to be placed directly on this surface if required. However, we would recommend that animals are placed on a thin layer of material such as 
a fleece mat. 

As air is not being circulated, the display will tend to under read by approximately 1 degree C in convection mode. 

 

In 

Fan Assisted 

mode, due to the flow of air around the inside of the cabinet, it is impossible to ensure a precise temperature  throughout, and this 

should be borne in mind during use.  Warmer and cooler areas are likely to exist, in particular, close to the cabinet side walls and towards the door 
respectively. 

 

On  initial  warm  up,  it  may  be  found  that  the  temperature  overshoots  the  set  point.    This  is  normal,  and  the  unit  should  be  left  to  stabilise  for  15 
minutes prior to use. 

 

The thermostat is accurate to approxi0.4°c and -0.2°c. 
 

Temperature Adjustment 

 

The set-point temperature is adjustable between factory set limits of 10 and 50°c in 0.1°c increments, and is factory set to 38.0°c 

 

To display the current set-

point value, press button ‘P’ momentarily.  The display will alternate between ‘SP’ and the set-point value.  To exit set-

point display press ‘P’ again, or if no buttons are pressed the display will revert to the current detected temperature automatically after 15 seconds. 

 

To adjust  the  set-

point value, press ‘P’ momentarily then use the 

UP

  and 

DOWN

 

arrow buttons to adjust the value higher or lower.  Pressing ‘P’ 

again, or leaving the display for 15 seconds, will exit the menu and save the current set-point value. 
 

Standby Mode 

 

If  required,  the  heating  elements  can  be  switched  off  leaving  the  fan  circulating  air.    To  put  the  device  in 

Standby

  mode  press  and  hold  th

e  ‘U’ 

button for approximately 3 seconds.  The display will shut down except for a Green or Red 

LED.  To resume operation press and hold the ‘U’ button 

for 3 seconds. 

 

If the power is interrupted at any time, the device will power up in the same condition it was in before the power loss. 
 

Maintenance 

 

The device is a sealed unit, so only the following items are user serviceable.  For repairs, please return to Peco Services Ltd. 

 

Cleaning -  

! Always disconnect the device from the mains power supply before cleaning it 

 

The  device  should  only  be  cleaned  with  mild  detergents  and  sterilising  solutions.    Under  no  circumstances  should solvents  of  any  kind  be  used.  
Use of such materials will damage the device. 

 

The  V1200  is  classified  as  IPX0.    This  means  that  it  is  not  waterproof  but  can  withstand  ingress  of  a  small  splash  of  water.    When  cleaning  the 
device, use a damp cloth and mild detergent.  Care should be taken to prevent contact with water or other liquids whilst in use. 

 

In the event of water or other liquids entering the device, it should be returned to the manufacturer for inspection/repair. 
 

Fan Filter 

 

The fan filter can be removed for cleaning.  It is located in a plastic frame clipped to the fan.  Remove the frame by carefully prising it away from the 
fan  using  your  fingers.    Clean  the  foam  filter  by  washing  in  soapy  water  and  leave  to  dry  in  air  at  room  temperature.    Replacement  filters  are 
available. 
 

Warranty 

The  V1200DT  is  guaranteed  against  faulty  parts  and  workmanship  for  a  period  of  12  months  from  date  of  purchase,  provided  it  is  used  in 
accordance with the above operating instructions. 
 
If a fault develops with your equipment, please either return to your supplier or directly to Peco Services Ltd at the address below. 
 
Peco Services Ltd reserves the right to repair or replace faulty goods at their discretion. 
 