5.2. Connection and Operation Procedure
1.) After checking with the rating label plug in to AC mains.
2.) Switch on the power supply and the LCD display should be on at the same time. The (CV)
icon should be shown on the display.
3.) Turn to current volume knob 5 to maximum clockwise
4.) If you do not require lower Current limiting value, otherwise do the preset the (CC) limiting
procedure.
5.) Set your desired output voltage and then turn off the output terminal by push button 6.
6.) Connect to your load positive to positive and negative to negative.
7.) Turn on the output terminal again and check if display shows (CV)
8.) If display shows (CC), either your preset current limiting value is too low or your load
requires more voltage and current. You need to re-access the voltage and current
requirement of your load and increase the voltage or current accordingly until (CV) appears.
5.3. Operations
5.3.1. Ground Connection
Depending on the application, the power supply output terminals can be grounded in any one of the
following grounding conditions:
1.) Negative ground - black (-) negative terminal is shorted with green GND terminal.
2.) Positive ground - red (+) positive terminal is shorted with green GND terminal.
3.) Floating ground - green terminal is not shorted with any of the output terminals.
Remarks:
When operating this power supply as a floating ground, high impedance leakage can exist between
the power supply circuitry and the chassis ground.
5.4. Basic Mode of Operation
The PeakTech 6095 / 6135 is designed to operate as a constant voltage source or as a constant
current source. Automatic crossover to either mode of operation occurs when the load condition
changes as following:
Constant Voltage (CV), Automatic crossover & Constant Current (CC)
The PeakTech 6095/6135 functions as a constant voltage source (CV) as long as the load current is
less than the preset current limiting value. When the load current is equal to or greater than the preset
current limiting value, the power supply will automatically cross over to the constant current mode,
voltage will drop, (CC) will show on the LCD display panel and it will operate as a constant current
source.
When the load current drops below the preset current limiting value, the supply returns to constant
voltage (CV) mode