5.1 Messurment in lm/ft² 

1.

Press  and  hold  the  "Range"  button  while  pressing  the  "ON"
button once to switch the device on and change to lm/ft² at the
same time.

2.

Select  the  measuring  range  with  the  RANGE-key  (3)  on  the
side of the unit

The three possible measurement ranges are: 

Range 

Display 

Resolution 

0 … 200 

FC 

0.1 FC 

2000 

x10 FC 

1 FC 

20000 

x100 FC 

10 FC 

Note: 1 lm/ft² is approx. 10.764 Lux 

After the desired measuring range is selected, you can start your 
measurements. If the indicating overflow "I" appears, the luminosity 
is higher than the selected measuring range. If necessary, switch 
the measuring range again until a measurement is displayed. 

To switch back to the LUX mode, turn the power off again, hold 
down the RANGE button again and turn the unit back on by 
simultaneously pressing the ON key. 

-16- 

Содержание 5160

Страница 1: ...PeakTech 5160 Bedienungsanleitung Operation Manual Lufttemperatur Feuchtigkeitsmessgerät Air Temperature Humidity Meter ...

Страница 2: ...oldering irons or guns away from the equipment Allow the equipment to stabilize at room temperature before taking up measurement important for exact measurements Replace the battery as soon as the battery indicator appears With a low battery the meter might produce false reading that can lead to electric shock and personal injury Fetch out the battery when the meter will not be used for long perio...

Страница 3: ...uipment Introduction This newly developed instrument measures the current humidity as well as the current air temperature installed a set of sensors for fast and accurate evaluation of the measurement results 2 Features 3 digit 10mm LCD display max 999 Automatic backlight Humidity measurement in R H Air temperature in C or F Fast response time of the sensor Rugged and compact housing design Data h...

Страница 4: ... C 10 30 RH 3 0 at 25 C 30 99 RH 1 C 2 6 F Measuring Rate 2 x second Low Battery Indication Symbol shown in the display if battery voltage is low Auto Power Off After 10 Minutes DATA HOLD Holds measurement value in display Power Supply 9V battery 6F22 or equivalent Operating Environment 10 C 40 C 99 RH Storage Environment 10 C 60 C 70 RH Size WxHxD 55 x 145 x 35mm Weight 158g 12 ...

Страница 5: ...escription 1 Temperature Humidity Sensor 2 ON Power switch 3 Temp Humidity button 4 HOLD button 5 Brightness sensor 6 Measured value display 7 Battery compartment rear 8 Thread for tripod rear 13 5 2 4 7 8 3 6 1 ...

Страница 6: ...gree 5 Operating Instruction 1 Press the ON button to turn on the unit 2 Select the desired measurement function Select the measurement function with the option key 3 on the device s side The three possible measurement functions are R H Relative Humidity C Temperature Measurement F Temperature Measurement After the desired measurement function is selected the device automatically switches from the...

Страница 7: ... vapor at room or ambient air 3 To freeze the current reading press the HOLD button The display stops updating the measured values and the HOLD symbol appears Automatic Power Off If the device is no longer used it will automatically turn off after 10 minutes Backlight This unit features an automatic backlight which activates the backlight when the light sensor on the front panel detects low light ...

Страница 8: ...Used up batteries are hazardous and must be given in the for this being supposed collective container NOTE 1 Keep the instrument dry 2 Keep the probes clean 3 Keep the instrument and battery out of reach of infant and child 4 When the symbol appears the battery is low and should be replaced immediately When you install battery ensure the polarity connections are correct If you will not use the ins...

Страница 9: ...of old batteries at a council collection point or return them to a local shop at no cost The disposal in domestic refuse is strictly forbidden according to the Battery Regulations You can return used batteries obtained from us at no charge at the address on the last side in this manual or by posting with sufficient stamps Contaminated batteries shall be marked with a symbol consisting of a crossed...

Страница 10: ...her only by written permission of the publisher This manual is according the latest technical knowing Technical alterations reserved We herewith confirm that the unit is calibrated by the factory according to the specifications as per the technical specifications We recommend to calibrate the unit again after one year PeakTech 08 2021 MP Ehr ...

Страница 11: ...PeakTech 5165 Bedienungsanleitung Operation Manual Digital Lux Meter ...

Страница 12: ...dering irons or guns away from the equipment Allow the equipment to stabilize at room temperature before taking up measurement important for exact measurements Replace the battery as soon as the battery indicator BAT appears With a low battery the meter might produce false reading that can lead to electric shock and personal injury Fetch out the battery when the meter will not be used for long per...

Страница 13: ... This newly developed instrument measures the current humidity as well as the current air temperature installed a set of sensors for fast and accurate evaluation of the measurement results 2 Features 3 1 2 digit 10mm LCD display max 1999 Automatic backlight Lux measurements up to 200 000 lux FC lm ft measurements up to 20000 FC Fast measurement rate of 1 5 measurements second Rugged and compact ho...

Страница 14: ...ly Accuracy 10k Lux 4 rdg 10k Lux 5 full Scale calibrated to standard incandescent lamp at color temperature 2 856 K Low Battery Indication Symbol shown in the display Auto Power Off After 10 Minutes DATA HOLD Holds measurement value in display Power Supply 9V battery 6F22 or equivalent Operating Environment 0 C 50 C 80 RH Storage Environment 10 C 50 C 70 RH Size WxHxD 55 x 155 x 35mm Weight 120g ...

Страница 15: ...iption 1 Light Sensor Photo Diode 2 ON Power switch 3 Range Switch 4 MAX MIN HOLD Switch 5 Brightness sensor backlight 6 Measured value display 7 Battery compartment rear 8 Thread for tripod rear 14 6 2 3 1 4 7 8 5 ...

Страница 16: ...Measurement unit of the luminosity in Lumen Square Feet lm ft 5 Operating Instruction 1 Press the ON button once to turn on the unit 2 Select the desired measurement range with the option key 3 on the device s side The three possible measurement ranges are Range Display resolution 0 2000 LUX 1 Lux 20000 x10 LUX 10 Lux 200000 x100 LUX 100 Lux After the desired measuring range is selected you can st...

Страница 17: ...0 200 FC 0 1 FC 2000 x10 FC 1 FC 20000 x100 FC 10 FC Note 1 lm ft is approx 10 764 Lux After the desired measuring range is selected you can start your measurements If the indicating overflow I appears the luminosity is higher than the selected measuring range If necessary switch the measuring range again until a measurement is displayed To switch back to the LUX mode turn the power off again hold...

Страница 18: ...ds the highest previous reading 3 Press the MAX MIN HOLD button again to switch to the minimum value mode The meter is now renewed only if the current measured value is lower than the previous reading 4 Press the button again to switch back to the normal measurement mode Automatic Power Off If the device is no longer used it automatically turns off after 10 minutes Backlight This unit features an ...

Страница 19: ...Used up batteries are hazardous and must be given in the for this being supposed collective container NOTE 1 Keep the instrument dry 2 Keep the probes clean 3 Keep the instrument and battery out of reach of infant and child 4 When the symbol appears the battery is low and should be replaced immediately When you install battery ensure the polarity connections are correct If you will not use the ins...

Страница 20: ...of old batteries at a council collection point or return them to a local shop at no cost The disposal in domestic refuse is strictly forbidden according to the Battery Regulations You can return used batteries obtained from us at no charge at the address on the last side in this manual or by posting with sufficient stamps Contaminated batteries shall be marked with a symbol consisting of a crossed...

Страница 21: ...e latest technical knowing Technical alterations reserved We herewith confirm that the unit is calibrated by the factory according to the specifications as per the technical specifications We recommend to calibrate the unit again after one year PeakTech 08 2021 MP Ehr PeakTech Prüf und Messtechnik GmbH Gerstenstieg 4 DE 22926 Ahrensburg Germany 49 0 4102 97398 80 49 0 4102 97398 99 info peaktech d...

Страница 22: ...PeakTech 5170 Bedienungsanleitung Operation Manual Digital Anemometer ...

Страница 23: ...ing irons or guns away from the equipment Allow the equipment to stabilize at room temperature before taking up measurement important for exact measurements Replace the battery as soon as the battery indicator appears With a low battery the meter might produce false reading that can lead to electric shock and personal injury Fetch out the battery when the meter will not be used for long period Per...

Страница 24: ...n meters second kilometers hour feet minute knots and miles per hour as well as the current air temperature with simultaneous display of these two measured values on the display 2 Features 3 5 6 digit 10mm LCD display max 5999 Automatic backlight illumination Multi line display for simultaneous value display Variety of meas units such as km h knots mph etc Air temperature display in C or F Maximum...

Страница 25: ...surement Range Air Temperature 10 45 C 14 113 F Messfolge 1 5 x Sekunde typisch Accuracy Windspeed 5 rdg Accuracy Air Temperature 2 C 3 6 F Low Battery Indication Symbol shown in the display Auto Power Off After 10 Minutes Power Supply 9V battery 6F22 or equivalent Operating Environment 10 C 45 C 80 RH Storage Environment 40 C 60 C 70 RH Size WxHxD 55 x 155 x 35mm Weight 120g 12 ...

Страница 26: ...escription 1 Anemometer probe 2 ON Power switch 3 MAX AVG Switch 4 Meas Unit Switch 5 Brightness sensor backlight 6 Dual meas value display 7 Battery compartment rear 8 Thread for tripod rear 13 3 6 1 2 4 7 8 5 ...

Страница 27: ...s mph Selected measurement unit C Measurement unit Air Temp in Celsius F Measurement unit Air Temp in Fahrenheit 5 Operating Instruction 1 Press the ON button once to turn on the unit 2 Select the measurement function with the unit selection button 4 on the device side The five possible modes are Range Unit Resolution Meters Second m s 0 1 m s Kilometers Hour Km h 0 1 km h Foot Minute ft min 1 ft ...

Страница 28: ...n to switch to the average mode The display indicates the average wind speed over the entire measurement 3 Press the button again to switch back to the normal measurement mode Automatic Power Off If the device is no longer used it automatically turns off after 10 minutes Backlight This unit features an automatic backlight which activates when the light sensor on the front panel detects low light c...

Страница 29: ...Used up batteries are hazardous and must be given in the for this being supposed collective container NOTE 1 Keep the instrument dry 2 Keep the probes clean 3 Keep the instrument and battery out of reach of infant and child 4 When the symbol appears the battery is low and should be replaced immediately When you install battery ensure the polarity connections are correct If you will not use the ins...

Страница 30: ...of old batteries at a council collection point or return them to a local shop at no cost The disposal in domestic refuse is strictly forbidden according to the Battery Regulations You can return used batteries obtained from us at no charge at the address on the last side in this manual or by posting with sufficient stamps Contaminated batteries shall be marked with a symbol consisting of a crossed...

Страница 31: ...her only by written permission of the publisher This manual is according the latest technical knowing Technical alterations reserved We herewith confirm that the unit is calibrated by the factory according to the specifications as per the technical specifications We recommend to calibrate the unit again after one year PeakTech 06 2020 MP EHR ...

Страница 32: ...PeakTech Prüf und Messtechnik GmbH Gerstenstieg 4 DE 22926 Ahrensburg Germany 49 0 4102 97398 80 49 0 4102 97398 99 info peaktech de www peaktech de ...

Страница 33: ...0 PeakTech 5175 Bedienungsanleitung Operation Manual Digital Schalpegelmessgerät Digital Sound Level Meter ...

Страница 34: ...g irons or guns away from the equipment Allow the equipment to stabilize at room temperature before taking up measurement important for exact measurements Replace the battery as soon as the battery indicator BAT appears With a low battery the meter might produce false reading that can lead to electric shock and personal injury Fetch out the battery when the meter will not be used for long period P...

Страница 35: ...is designed for the measurement of sound sources in decibels by A weighting dBA which is based on the human ear 2 Features 3 1 2 digit 10mm LCD display max 1999 Fast 125 ms and Slow 1 s measurement sequence Sound Level dBA rate Maximun value measurement Minimum value measurement Acoustic and visual alarm function in excess of a user defined sound level in dB Rugged and compact housing design Suita...

Страница 36: ...t 125 ms Slow 1 s Accuracy 1 5 dB Microphone in Electret Condenser Self calibration time 10 Sek every power on Low Battery Indication Symbol shown in the display Auto Power Off Off After 10 Minutes Power Supply 9V battery 6F22 or equivalent Operating Environment 0 C 40 C 10 80 RH Storage Environment 10 C 60 C 10 70 RH Size WxHxD 55 x 150 x 35mm Weight 160g ...

Страница 37: ...l Description 1 Wind shield microphone 2 ON OFF Power switch 3 Fast Slow Switch 4 MAX MIN Switch 5 LED for Alarm function 6 Meas value display 7 Battery compartment rear 8 Thread for tripod rear 1 2 4 5 7 8 3 6 ...

Страница 38: ... unit Decibel in A weighting 5 Operating Instruction 1 Press the ON button once to turn on the unit 2 Select the measurement functions with the selection buttons on the device side The four possible functions are Function Description Fast Fast time weighting with 125 ms which also shows short sound bursts Slow Slow time weighting with 1 second which shows an average sound value over the second Sui...

Страница 39: ...red value is less than the lowest previous reading 5 Press the button 4 again to switch back to the normal measurement mode Automatic Power Off If the device is not used it automatically turns off after 10 minutes For continuous measurements the automatic shutdown can be disabled 1 Turn off the unit if necessary with the ON button 2 2 Turn on the unit again with the ON button 2 and hold the button...

Страница 40: ... level will be detect over the set value the unit will alarm LED flash 01 alarm mode or only LED flash 02 alarm mode Alarm will stop continue 15 seconds After stop 5 seconds If unit detect the sound level over set value the unit will alarm again In this mode the unit will not automatic power off Press MAX MIN key again cancel alarm mode See Fig 5 ...

Страница 41: ...ncrease the alarm values of 0 1dB Continuous pressing increases the value in 1dB steps Press MAX MIN key to confirm 3 Select the alarm mode To select the alarm mode press the MAX MIN button 2 times then press and hold the MAX MIN button until the display shows a stable value Then press the FAST SLOW button to select the alarm mode 01 Alarm mode audible and visual signal 02 Alarm mode only visual s...

Страница 42: ... Used up batteries are hazardous and must be given in the for this being supposed collective container NOTE 1 Keep the instrument dry 2 Keep the probes clean 3 Keep the instrument and battery out of reach of infant and child 4 When the symbol appears the battery is low and should be replaced immediately When you install battery ensure the polarity connections are correct If you will not use the in...

Страница 43: ...se of old batteries at a council collection point or return them to a local shop at no cost The disposal in domestic refuse is strictly forbidden according to the Battery Regulations You can return used batteries obtained from us at no charge at the address on the last side in this manual or by posting with sufficient stamps Contaminated batteries shall be marked with a symbol consisting of a cros...

Страница 44: ...the latest technical knowing Technical alterations reserved We herewith confirm that the unit is calibrated by the factory according to the specifications as per the technical specifications We recommend to calibrate the unit again after one year PeakTech 01 2021 MP HR PeakTech Prüf und Messtechnik GmbH Gerstenstieg 4 DE 22926 Ahrensburg Germany 49 0 4102 97398 80 49 0 4102 97398 99 info peaktech ...

Отзывы: