The meter is designed to withstand the stated max
voltages. If it is not possible to exclude without that
impulses, transients, disturbance or for other reasons,
these voltages are exceeded a suitable presale (10:1)
must be used.
Replace a defective fuse only with a fuse of the original
rating. Never short-circuit fuse or fuse holding.
Disconnect test leads or probe from the measuring circuit
before switching modes or functions.
Do not conduct voltage measurements with the test leads
connected to the µAmA/A- and COM-terminal of the
equipment.
The 10A-range is protected by fuse 10A/1000V.
To avoid electric shock, disconnect power to the unit under
test and discharge all capacitors before taking any
resistance measurements.
Do not conduct current measurements with the leads
connected to the V/Ω-terminals of the equipment.
Check test leads and probes for faulty insulation or bare
wires before connection to the equipment.
Please use only 4mm-safety test leads to ensure
immaculate function.
To avoid electric shock, do not operate this product in wet
or damp conditions. Conduct measuring works only in dry
clothing and rubber shoes, i. e. on isolating mats.
Never touch the tips of the test leads or probe.
Comply with the warning labels and other info on the
equipment.
The measurement instrument is not to be to operated
unattended.
Always start with the highest measuring range when
measuring unknown values.
Do not subject the equipment to direct sunlight or extreme
temperatures, humidity or dampness.
-32-