6.6. Inductance measurements
Warning !
To avoid electric shock, disconnect power to the unit under test
and discharge all capacitors before taking any capacitance
measurements. Remove the batteries and unplug the line cords.
1. Set the function switch to the desired Inductance-range
(600µH, 6mH, 60mH, 600mH, 6H)
2. Insert
the red test lead into the “L”-terminal and the black
test
lead into the “L/A”-terminal.
3.
Press “REL”-button to set the display to zero.
4. Connect the test leads crossly to the two ends of inductor
and read the measurement value from the LCD.
Notes:
1. If inductance exceeds the selected
range “OL” will be
displayed on LCD and you shall increase the range by on
step.
2. The inductance value measured for identical inductor may
be different if there is different impedance.
3. Avoid measuring small inductor in high range, or the
accuracy is not guaranteed.
-62-