37. Technical Specifications
Unless otherwise specified, the technical specifications applied are applicable to the probe with the
attenuation switch setting 10X and this digital oscilloscope. Only if the oscilloscope fulfills the following two
conditions at first, can these specification standards be reached.
*
This instrument should run for more than 30 minutes continuously under the specified operating
temperature.
*
If the change range of the operating temperature is up to or exceeds 5°C, open the system function menu
and execute the “Auto-calibration” procedure (see “7. How to Implement Slef-calibration” on page 102).
All specification standards can be fulfilled, except one(s) marked with the word “Typical”.
Performance Characteristics
Instruction
Bandwidth
P 1240
60 MHz
P 1245
100 MHz
P 1255
100 MHz
P 1260
200 MHz
Channel
2 + 1 (External)
Acquisition
Mode
Normal, Peak detect, Averaging
Sample rate
(real time)
P 1240
Dual CH
250 MSa/s
Single CH
500 MSa/s
P 1245
Dual CH
500 MSa/s
Single CH
1 GSa/s
P 1255
Dual CH
1 GSa/s
Single CH
2 GSa/s
P 1260
Dual CH
1 GSa/s
Single CH
2 GSa/s
Input
Input coupling
DC, AC , Ground
Input impedance
1M
Ω
±2%, in parallel with 10pF±5pF
Probe
attenuation
factor
1X, 10X, 100X, 1000X
Max. input voltage
400V (PK-PK) (DC + AC PK-PK)
Bandwidth limit
20MHz, full bandwidth
Channel–channel
isolation
50Hz: 100 : 1
10MHz: 40 : 1
Time delay between
channel(typical)
150ps