The description of the
cursor measurement menu
is shown as the following table:
Function Menu
Setting
Description
Type
OFF
Voltage
Time
Switch off the cursor measurement.
Display the voltage measurement cursor and menu.
Display the time measurement cursor and menu.
Source
CH1
CH2
Display the channel generating the waveform to which the
cursor measurement will be applied.
Delta
Read the difference between cursors.
Cursor 1
Read the position of Cursor 1 (the Freq. is read with
reference to the horizontal trigger position and the Vamp is
to the ground point).
Cursor 2
Read the position of Cursor 2 (the Freq. is read with
reference to the horizontal trigger position and the Vamp is
to the ground point).
When carrying out the cursor measurement, the position of Cursor 1 can be adjusted with the
CURSOR1
(VERTICAL POSITION)
knob of Channel 1, and that of Cursor 2 can be adjusted with the
CURSOR2
(VERTICAL POSITION)
knob
of Channel 2.
Perform the following operation steps for the voltage cursor measurement of the channel CH1:
1.
Press
CURSOR
and recall the
CURS MEAS
menu.
2.
Press the
F1
menu selection button to choose
Voltage
for Type, with two purple dotted lines displayed
along the horizontal direction of the screen, which indicating
CURSOR1
and
CURSOR2
.
3.
Press the
F2
menu selection button and choose
CH1
for
Source
.
4.
Adjust the positions of
CURSOR1
and
CURSOR2
according to the measured waveform, with the
absolute value of the voltage amplitude difference between Cursor 1 and Cursor 2 displayed in the
increment window. The current position of Cursor1 is displayed in the Cursor1 window and that of
Cursor2 is displayed in the Cursor2 window (see Fig.68).
Fig. 68
Waveform of Voltage Cursor Measurement
-181-