MU-Thermocouple1 CAN – User Manual
15
5
Technical Specifications
Measuring modules
Supported thermocouple types
(IEC 584)
J (-210 - +1121 °C)
K (-200 - +1370 °C)
T (-200 - +400 °C)
(according to equipment)
Connectors
Thermocouple socket with standard size (DIN EN
50212)
Reference sensors
4 (1 per measuring module having 2 sockets)
Measuring category
CAT I (only electric circuits that are not
connected to the mains)
Measurement accuracy
±0,2 %
Accuracy of the reference
sensors
±0,5 K at +25 °C
±1 K at 0 - +70 °C
±2 K at -20 - +85 °C
±3 K at -40 - +125 °C
Controller module
Microcontroller Philips/NXP
LPC2364
Temperature resolution at CAN
transmission
1/16 K
CAN
Specification ISO
11898-2
High-speed CAN (up to 1 MBit/s)
2.0A (standard format) and 2.0B (extended
format)
Transceiver
NXP (Philips) TJA1040T
Connection
D-Sub (m), 9 pins, pin assignment according to
CiA recommendation 102 DS
Optional 5-Volt supply at pin 1 for external
devices (e.g. bus converters), max. 100 mA
Galvanic isolation up to 500 V
Continued on the next page