Probe compensation continued…
When prompted, attach a small length of
tinned copper wire between the two test
probes. Now let the leads rest on a non-
conductive surface, try not to touch them
during the compensation procedure.
When you’re happy that a suitable good link is present
between the probes, briefly press the
enter-off
button.
After a short time, the instrument will ask
you to remove the link from the probes.
At that point, remove the link, rest the probes on a non-
conducting surface and don’t touch them. When you’ve
done that you can briefly press
enter-off
.
If the LCR45 is happy with the measurements that it has taken during the
probe compensation procedure, you will see the following message:
Probe properties
stored.
Please note that probe compensation is particularly important when
analysing low value inductors, capacitors and resistors.
Short the probes
then press enter
Open the probes
then press enter