Appendix C – Technical Specifications
Parameter
Min
Typ
Max
Note
Resistance
range
1Ω
2MΩ
resolution
0.3 Ω
0.6Ω
accuracy
Typically ±1.0% ±1.2Ω
1,2,6
Capacitance
range
0.5pF
10,000µF
resolution
0.2pF
0.5pF
accuracy
Typically ±1.5% ±1.0pF
1,2,5
Inductance
range
1µH
10H
resolution
0.4µH
0.8µH
accuracy
Typically ±1.5% ±1.6µH
1,2,4
Peak test voltage (across O/C)
-1.05V
+1.05V
Peak test current (thru S/C)
-5.0mA
+5.0mA
Test frequency
accuracy
1kHz
-1.5%
±1%
+1.5%
14.925kHz
-1.5%
±1%
+1.5%
200kHz
-1.5%
±1%
+1.5%
Sine purity
Typically -60dB 3
rd
harmonic
Operating temperature range
10°C
40°C
3
Battery operating voltage
8.5V
13V
Notes:
1.
Within 12 months of factory calibration. Please contact us if you
require a full re-calibration and/or certification of traceable
calibration.
2.
Specified at temperatures between 15°C and 30°C.
3.
Subject to acceptable LCD visibility.
4.
For inductances between 100µH and 100mH.
5.
For capacitances between 200pF and 500nF.
6.
For resistances between 10Ω and 1MΩ.