User Manual
Page 3
Peak Scientific – Rev 9A – 16/09/15
Stage 1
20
Stage 2
21
Service Indication Reset
Service Plans
Cleaning
Run Time Alarm
22
Stage 3
Trouble Shooting
23
Страница 1: ...ABN2ZA User Manual 2015 Peak Scientific Rev 9A 16 09 15...
Страница 2: ...ctrical Requirements 8 General 8 Unpacking 9 Installation 10 Generator Environment 10 General Dimensions 10 Removal of Transit Brackets 11 Unit Controls 12 Rear Connections 13 Drain Connection 13 Elec...
Страница 3: ...ZA User Manual Page 3 Peak Scientific Rev 9A 16 09 15 Stage 1 20 Stage 2 21 Service Indication Reset 21 Service Plans 21 Cleaning 21 Run Time Alarm 22 Stage 1 22 Stage 2 22 Stage 3 22 Trouble Shooting...
Страница 4: ...date Liam Couttie 16 09 2015 How to use this Manual This manual is intended for end users and has been written so that it can either be read as a step by step guide to installation and usage or as a r...
Страница 5: ...lity Re heat technology improves membrane performance and reliability Robust control system improves safety and reliability of units With the ABN2ZA based on proven technology it selectively removes o...
Страница 6: ...out prejudice to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and or servicing incorrect repair and or servicing use of wrong materials and or...
Страница 7: ...egulations The attention of UK users is drawn to the Health and Safety at Work Act 1974 and the Institute of Electrical Engineers regulations A WARNING notice denotes a hazard It calls attention to an...
Страница 8: ...n 7 6 bar 0 85 cfm 110 psi Exhaust maximum flow 8 L min 4 2 bar 0 28 cfm 60 psi Particles 0 01 m Phthalates NONE Suspended liquids NONE Gas outlets 3 x BSPP Drain outlet 1 x BSPP Pressure gauges 3 Ele...
Страница 9: ...to the carrier and Peak Scientific or the Peak Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate It will require two people to remove the uni...
Страница 10: ...ions Note should also be taken to the proximity of Air Conditioning outlets These can sometimes give rise to pockets of air with high relative humidity Operation of the unit within such a pocket could...
Страница 11: ...it remove the side covers from the cabinet ensuring that earth cables are disconnected 2 Remove the four screws with a 3 Pozidrive Screwdriver See figure 2 3 Slide the transit bracket out from under t...
Страница 12: ...12 Peak Scientific Rev 9A 16 09 15 Unit Controls Figure 4 Unit controls Figure 5 PLC display Left Hand Side Right Hand Side Power switch Mute button Alarm LED Service LED Pressure gauges Curtain Sour...
Страница 13: ...enerator See Figure 6 Drain Connection Fit the 6mm push fit fitting item 2 from fittings kit to the drain port located on the output panel see Figure 6 Tighten using a 16mm or 5 8 spanner Use the 6mm...
Страница 14: ...Generator can accommodate transients between 207VAC and 253VAC However running continuously at voltages less than 220V is not recommended and extended periods at these extremes can have a detrimental...
Страница 15: ...commend fitting a transformer This can be ordered directly from Peak Scientific with the order number being as below Table 1 Additional Transformer On completing the voltage check the Generator should...
Страница 16: ...ications Once these pressures are reached the compressors will continue to run until the internal tank upper pressure limits 125psi for Nitrogen system and 130psi for Dry Air system are reached and th...
Страница 17: ...are thoroughly checked for being leak tight Even the slightest leak in the gas supply between the Generator and the Mass Spectrometer can lead to a reduction in efficiency or possible alarm of the Gen...
Страница 18: ...L ADDRESS CITY TOWN GENERATOR SERIAL NUMBER POSTCODE COUNTRY MODEL TYPE TELEPHONE INSTALLATION DATE DD MM YYYY Important Please Note You have 1 month to register your Peak Scientific product from the...
Страница 19: ...e requirements of the Mass Spectrometer If the Mass Spectrometers requirement for gas stops the Generator will also stop once it has reached its upper set limit in the internal storage tanks If the de...
Страница 20: ...6 5529 2 Table 2 Service schedule Compressors can be re fitted as an alternative to replacement up to a maximum of 3 times This is a more cost effective solution although a degree of technical experti...
Страница 21: ...ormed by the Peak Service Engineer or trained service representative that completes the service operation Service Plans Peak Scientific offer two service plans The Complete Service Plan specifically d...
Страница 22: ...constantly If the Generator returns to a cycling operation after the 8 hours then the alarm will cancel the Alarm LED indicator will go out and the buzzer will stop sounding Stage 2 If the control sy...
Страница 23: ...cation section of this manual for further information Yellow LED on front panel is flashing A compressor s is overdue for service Contact your service provider urgently Refer to Service Indication sec...
Страница 24: ...hongzhi Road Pudong New Area Shanghai 201203 China Peak Scientific Mexico Solon 352 Col Los Morales Polanco 11530 Mexico D F Mexico Peak Scientific Australia PO Box 65 Belmont Victoria 3216 Australia...