Page 16
Installation
Generator Environment
The generator is designed for indoor use only. Storing of corrosive substances near
the generator or its compressed air supply is not recommended. Saline salt solution
spillages around the generator or its compressed air supply should be cleaned up
immediately. It should be installed adjacent to the application(s) it is supplying. If this
is not convenient then the unit can be sited elsewhere, however, consideration should
be made of the lengths of pipe runs as pressure drops can result from extended runs of
pipe.
Performance of the generator (like all sophisticated equipment) is affected by ambient
conditions. Note should also be taken to the proximity of Air Conditioning outlets.
These can sometimes give rise to “pockets” of air with high relative humidity. Operation
of the unit within such a pocket could adversely affect its performance. Consideration
should also be given to the air flow around the unit. It is recommended that an air gap
of 75mm (3”) should be maintained down both sides and at the rear of the unit. Please
refer to the drawing below for the general dimensions of the unit.
Please ensure generator is operated in a well ventilated environment and is positioned
so as to permit easy electrical disconnection if required.
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
35 °C (95 °F)