*Failure to change the DI cartridge can lead to cell damage and will invalidate your warranty

Step 2:

 Open the top panel by 

lifting gently.

Simple user maintenance guide

De-ionizer cartridge replacement

When to replace?

 Every 12 months from original installation.

Step 3: 

The de-ionizer cartridge is 

located in the water tank and can 
be accessed by opening the top 
panel.

Step 4:

 The de-ionizer cartridge 

can be removed by turning the 
cartridge counter-clockwise.

Step 5:

 Remove the de-ionizer 

cartridge and replace with a fresh 
cartridge.

Step 6: 

Fix replacement 

de-ionizer cartridge in place by 
turning clockwise.

Step 7:

 Close the top panel. The 

de-ionizer cartridge will need to be 
changed after 12 months.

Step 1: 

Press and hold the power 

button for 5 seconds to shutdown 
and depressurize. Now turn off the 
main power switch at the back of 
the unit. 

Step 8: 

Power up the generator using the main power switch on 

the rear of the unit. Once diagnostics are complete, the LEDs will 
be solid green and the generator will be ready to deliver gas to the 
application. Press the front button to supply gas to the application.

Please ensure

 Please ensure that you change the de-ionizer (DI) cartridge as soon as you unpack the 

replacement. The DI cartridge needs to be wet when installed.  If the DI cartridge has dried out it will
affect generator performance. To reactivate a dry DI cartridge, submerge it upside down in deionized
water so the air inside escapes, then fit as described in the following steps.