3)Insert the " Testing bottle " to the bottom of
" Container " ( 3-2, Fig. 1 ) completely.
4)Before the measurement, it should envelop in the
" Cover " ( 3-1, Fig. 1 ) completely.
Remark :
Before the measurement, it should
keep the outside of Testing Bottle
under the dry condition and without
existing any dust.
4-2 Measurement
1)Fill the measurement liquid into the " Testing bottle "
( 3-11, 3-12, Fig. 1 ) .
Attention :
Model 0601
It should fill the
10 mL
measurement
liquid capacity
until its level
up to the " level
marker " of the
" Testing bottle "
2)Insert the " Testing bottle " to the bottom of
" Container " ( 3-2, Fig. 1 ) completely.
3)Before the measurement, it should envelop in the
" Cover " ( 3-1, Fig. 1 ) completely.
4)Power ON the meter by pressing the " Power Button "
( 3-6, Fig. 1 ) once.
5)Press " TEST Button " ( 3-5, Fig. 1 ) once, the display
will show the text " tESt " ( TEST )", following the "
tESt " will flash approx. 10 seconds , then the TURBIDITY
value will present on the display along with the " ntu "
unit.
5