16  Dissolved Oxygen Measurement 

PCE-WO2  10  pocket  dissolved  oxygen  tester  is suitable for measuring  the  water,  wastewater, 
brine and other liquids. If the sample is belong to the seawater or other water containing large 
amounts of salt, please setting the salinity coefficient before measurement. Some gas and steam 
such as chloride, sulfur  
dioxide, sulfureted hydrogen, ammonium, carbon dioxide and iodin can permeate the membrane 
via diffusion. So their existence will influence the measurement of dissolved oxygen. If the sample 
contains  the  solvent,  grease,  sulfide  and  alga,  the  membrane  on  the  probe  will  be  blocked, 
damaged or eroded. 
 

 

Connect the dissolved oxygen probe to tester and wait for 15 minutes to polarize the 
probe. 

 

If necessary, to set the barometric pressure and salinity coefficient in the setup menu 
(Refer to page 5). 

 

Immerse the probe in the sample solution, make sure the temperature sensor on the 
probe is fully immersed. 

 

Stir the probe gently. Record the measured value when the reading is stable. 

 

 

 

17  Auto-Hold 

PCE-WO2  10  pocket  dissolved  oxygen  tester  contains  an  Auto-Hold  function.  If  enabled,  the 
tester will automatically sense a stable reading and lock the  
measurements. The 

 icon appears on the display. If disabled, press the 

Hold

 key, the tester 

will immediately lock the displayed value. Press the 

Meas

 key to resume measuring. 

 

 

 

 

Temperature  
sensor 

Содержание PCE-WO2 10

Страница 1: ...0 Oxygen Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Страница 2: ...o Use 6 11 Setup Menu 6 11 1 Setting the default option 7 12 Setting the salinity coefficient 8 13 Setting the ambient pressure 8 14 DO Calibration in Saturation Mode 9 14 1 Single point calibration 1...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...ure compensation 0 40 C 32 104 F automatic Air pressure correction 60 0 112 5 kPa 450 850 mmHg Salinity correction 0 35 g l Sensor type polarography Environmental conditions 0 60 C 32 140 F max 80 r H...

Страница 5: ...lay PCE WO2 10 pocket dissolved oxygen tester is equipped with an easy to read LCD display that used to show the measured values and mode icons The following table describes the function of each icon...

Страница 6: ...hes the tester ON OFF Locks the measured value press the key again to resume measuring Exits the calibration or setting and returns to the measurement mode CAL Press the key to start the calibration P...

Страница 7: ...cap onto the probe excess electrolyte solution will drain out Be sure the cathode of probe makes contact with membrane cap the electrolyte solution in membrane cap should be without an air bubble 8 C...

Страница 8: ...uirements MENU DESCRIPTION OPTIONS DEFAULT Set the salinity coefficient of sample Range 0 35ppt 0 0ppt Set the barometric pressure coefficient Range 450 850mmHg 760mmHg Set the number of calibration p...

Страница 9: ...the steps above until the tester returns to the measurement mode Setting is completed During the setting process press the Meas key the tester will exit the setup menu and return to the measurement m...

Страница 10: ...in the unit mmHg The following table will help to convert your ambient pressure into the required unit mmHg on the basis of your current height ALTITUDE m kPa mmHg ALTITUDE m kPa mmHg 0 101 3 760 160...

Страница 11: ...e into the air saturated water for 15 minutes Press the Enter key Wait for the reading to stabilize the tester automatically shows END Single point calibration is completed 14 2 2 points calibration M...

Страница 12: ...ss and hold the Cal key for 3 seconds to enter the setup menu Press the Enter key until the tester shows C CAL or F CAL Press the Cal key to enter the temperature calibration mode Press the Cal key ag...

Страница 13: ...the membrane on the probe will be blocked damaged or eroded Connect the dissolved oxygen probe to tester and wait for 15 minutes to polarize the probe If necessary to set the barometric pressure and...

Страница 14: ...ECTIVE ACTION DO probe does not connect to tester Check the connector of probe Measured value is out of range Check the DO membrane whether clogged dirty or broken Electrolyte solution is depleted Ref...

Страница 15: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Страница 16: ...0583 975 114 Fax 49 0 2903 976 99 9971 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce instruments com info industrial needs com info pce italia it www pce instruments com deutsch www pce instrumen...

Отзывы: