© PCE Instruments
47
5.8.4
Brightness
In this tab, the display brightness can be adjusted from 10 % to 100 %. Automatic dimming can
also be set. After a set time, the display w ill be dimmed to a low er brightness in order to save
energy. Pressing any key w ill set the brightness back to its original value.
5.8.5
Language
This menu is used to sw itch betw een different menu languages. The available languages are
English, German, French, Spanish, Italian, Dutch, Portuguese, Turkish, Polish, Russian, Chinese
and Japanese.
5.8.6
Auto power off
This option enables the auto pow er off function. The available time periods are 1 minute, 5
minutes and 15 minutes. After the set time period has elapsed, the device will automatically turn
off and pressing any key w ill reset the timer. It is also possible to disable the auto pow er off
function.
5.8.7
Reset
This menu is used to reset the device to its factory settings. Device settings are separate from
the calibration and each can be reset individually by selecting the corresponding menu item.
A reset of the device settings w ill load default values for the measurement parameters and any
remaining menu options. A calibration w hich may have been performed beforehand is retained.
In order to prevent an inadvertent reset of the calibration, a code is required to enter this menu.
The code is the same as for the calibration itself:
1402
.
Fig. 13 Code request
ATTENTION:
When the calibration is reset, the previously performed and saved calibration w ill be deleted and
a default calibration for the supplied sensor w ill be selected. It is recommended to perform a
calibration after the reset.
The reset must be confirmed via the subsequent confirmation dialogue. The device w ill
automatically restart after the reset.
Fig. 14 Confirmation dialogue
E
n
g
lis
h