- 9 -
Select “correct white” in control of the window and Put the testing aperture on the
white tabula, press “test”. Then the value of XYZ will appear in the dialog window
as follows:
If the value needs to be amended, fill the needed values in the dialog window. Click
“modificate” for saving the amending or “cancel” for deleting the amending.
Note: When doing correct white calibration. Please follow the above steps strictly. If
the absolute values of X
、
Y
、
Z are bigger than 0.1, correct white is necessary to be
done.
6.5 Enter sampling