background image

Manual 

Specification 

Measurement range

  

 

Ra

  

0.05 ... 10 µm / 1 ... 400 µinch

  

Rz

  

0.02 ... 100 µm / 0.78 ... 40 µinch

  

Measurement accuracy

  

 

General

  

±10 %

  

Fluctuation on display

  

Max. 6 %

  

Resolution

  

 

measurement range < 10 µ

  

0.001 µ

  

measurement range < 100 µ

  

0.01 µ

  

measurement range ≥ 100 µm

  

0.1 µ

  

Sensor

  

Measuring type

  

Induction principle

  

Radius of sensor

  

10 µm

  

Material of sensor

  

Diamond

  

Measuring force of sensor

  

16 mN (1.6gf)

  

Sensor angle

  

90 °

  

Vertical radius of probe head

  

48 mm

  

Measurement speed

  

 

Sampling length: 0.25

  

Vt = 0.135 mm/s

  

Sampling length: 0.8

  

Vt = 0.5 mm/s

  

Sampling length: 2.5

  

Vt = 1 mm/s

  

Returning

  

Vt = 1 mm/s

  

General Technical Specifications

  

Maximum hub

  

2.5 mm / 0.5 inch

  

Limit of frequency length

  

0.25 mm / 0.8 mm / 2.5 mm optical

  

Units

  

Rz, Ra

  

Display

  

4-digit LCD display, with background lights

  

Maximum surrounding temperature

  

0 ... +50 °C

  

Maximum surrounding air humidity

  

< 80%

  

Diameters

  

140 x 52 x 48 mm

  

Weight

  

280 g

  

 
 

 

Содержание PCE-RT 10

Страница 1: ...ts UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce instruments com english www pce instruments com Manual Surface Texture Meter PCE RT 10 Version 1 0 Date of creation 21 04 2016 Date of last change 04 05 2016 ...

Страница 2: ...cedure 6 4 3 1 Preparations 6 4 3 2 Perform measurements 7 4 3 3 Set the measurement length 7 4 3 4 Calibration 7 4 3 5 Set Cutoff length 7 4 4 Notes 7 4 4 1 General Notes 7 4 4 2 Standards 7 4 4 3 Travel length 8 4 4 4 Recommended length of the cutoff frequency 8 5 Battery replacement 8 6 Disposal 9 7 Contact 9 7 1 PCE Instruments UK 9 7 2 PCE Americas 9 ...

Страница 3: ...it will turn itself off automatically 2 Safety notes Please read this manual carefully and completely before you use the device for the first time There is no warranty of damages or injuries caused by non observance of the manual The instrument may only be used as described in this manual If it is not used in the described way it may cause injuries or break Only use the device in the prescribed am...

Страница 4: ... of sensor Diamond Measuring force of sensor 16 mN 1 6gf Sensor angle 90 Vertical radius of probe head 48 mm Measurement speed Sampling length 0 25 Vt 0 135 mm s Sampling length 0 8 Vt 0 5 mm s Sampling length 2 5 Vt 1 mm s Returning Vt 1 mm s General Technical Specifications Maximum hub 2 5 mm 0 5 inch Limit of frequency length 0 25 mm 0 8 mm 2 5 mm optical Units Rz Ra Display 4 digit LCD display...

Страница 5: ... Chosen parameter 5 Parameter button Up 6 Calibration button 7 Power on off 8 Battery level 9 Measurement unit 10 Cutoff frequency 11 Start button 12 Cutoff frequency button 13 µm µinch button Down 4 2 Delivery content 1 x Surface texture meter PCE RT 10 1 x Calibration plate 1 x User s manual 1 x Screwdriver ...

Страница 6: ...ess the cutoff frequency button To set the correct length of the cutoff frequency refer to the table Recommended length of the cutoff frequency within this manual 5 Press the µm µinch button to check the correctness of the measuring units 6 Clean the surface you want to measure in order to avoid faulty measuring results 7 The device needs to be placed correctly on the surface This is shown in the ...

Страница 7: ... shows LEN You can now choose length 1 and 2 by pressing the Up or Down button The settings can be saved by pressing the Start button 4 3 4 Calibration To calibrate the device press the Calibration button CAL will show up in the display Now place the device on the calibration plate and press the Start button The display will show a result after measurement has been made You can now adjust value ac...

Страница 8: ...3 10 0 8 0 63 1 25 3 2 6 3 0 32 0 63 1 6 3 2 0 25 0 32 2 25 1 6 0 25 0 20 0 25 0 16 0 20 1 0 1 25 0 8 1 0 0 125 0 16 0 1 0 125 0 08 0 1 0 63 0 8 0 5 0 63 0 4 0 5 0 063 0 08 0 05 0 063 0 04 0 05 0 32 0 4 0 25 0 32 0 2 0 25 0 032 0 04 0 025 0 032 0 02 0 025 0 16 0 2 0 125 0 16 0 1 0 125 5 Battery replacement The batteries need to be replaced when indicated on the display Open the battery lid and rep...

Страница 9: ...e use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 7 Contact If you have any questions about our range of products or measuring instruments please contact PCE Instruments 7 1 PCE Instruments UK By post PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign Way Southampton Hampshire Unite...

Отзывы: