© PCE Instruments
2
2
Specifications
Measuring range
0 ... 200 N
Calibration
0.1 N
Pressure calibration
1 Mpa
Load cell
integrated load cell with M6 connection
Measuring range
1 ... 100 % f. s.
Measurement
accuracy
±0.5 %
Units
n, kg, lb
Display
LCD
Operating
temperature
+10 ... +30 °C
Relative humidity
15 ... 80 %
Working conditions
The device must not be located near sources of vibration or corrosive
substances
Weight
1 kg / 2.2 lb
3
Important information before use
•
Wear protective mask and gloves.
•
Do not use damaged or bent fixtures.
•
If “Err-1” appears on the LCD, it indicates that the test load the sensor measures is 110
% larger than the rated load. Decrease the load immediately. Make sure the load
applied is smaller than 105 % the rated load after the meter is restarted.
•
Make sure the force gauge is not overloaded. Overload may damage the sensor.
•
Do not use sharp tools to press buttons.
•
Keep the meter away from liquids and store it in a cool, dry place and away from
sources of vibration.
•
Only use the charger delivered with the force gauge for charging.
•
Connect the meter to the computer in line with the instructions to avoid problems.
•
Only connect the charger to an intact socket as damaged sockets can cause fire.
•
Make sure the charger is securely inserted into a socket if the product needs to be
charged. Looseness may cause short circuit, resulting in electric shock or fire.
•
Remove the charger after charging to avoid accidents.
•
Do not touch the power adaptor with wet hands.
•
Do not bend or twist the test head. A measurement may only be made perpendicularly
to the meter.