Specifications 

Measuring range 

0 ... 200 N 

Calibration 

0.1 N 

Pressure calibration 

1 Mpa 

Load cell 

integrated load cell with M6 connection 

Measuring range 

1 ... 100 % f. s. 

Measurement 
accuracy 

±0.5 %  

Units 

n, kg, lb 

Display 

LCD 

Operating 
temperature 

+10 ... +30 °C 

Relative humidity 

15 ... 80 % 

Working conditions 

The device must not be located near sources of vibration or corrosive 
substances 

Weight 

1 kg / 2.2 lb 

 

Important information before use 

 

Wear protective mask and gloves. 

 

Do not use damaged or bent fixtures. 

 

If “Err-1” appears on the LCD, it indicates that the test load the sensor measures is 110 
%  larger  than  the  rated  load.  Decrease  the  load  immediately.  Make  sure  the  load 
applied is smaller than 105 % the rated load after the meter is restarted. 

 

Make sure the force gauge is not overloaded. Overload may damage the sensor. 

 

Do not use sharp tools to press buttons. 

 

Keep  the  meter  away  from  liquids  and  store  it  in  a  cool,  dry  place  and  away  from 
sources of vibration. 

 

Only use the charger delivered with the force gauge for charging. 

 

Connect the meter to the computer in line with the instructions to avoid problems. 

 

Only connect the charger to an intact socket as damaged sockets can cause fire. 

 

Make  sure  the  charger  is  securely  inserted  into  a  socket  if  the  product  needs  to  be 
charged. Looseness may cause short circuit, resulting in electric shock or fire. 

 

Remove the charger after charging to avoid accidents. 

 

Do not touch the power adaptor with wet hands. 

 

Do not bend or twist the test head. A measurement may only be made perpendicularly 
to the meter. 
 

 

 

Содержание PCE-PTR 200N

Страница 1: ...PCE PTR 200N Dynamometer User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change 2...

Страница 2: ...ption 4 7 1 Keys and adaptor 4 7 2 Dimensions 5 7 3 Circuit diagram of 3 5 mm jack 5 7 4 Display 6 7 5 Keys 7 8 Operation 8 9 Options 10 10 Alarm 11 11 Calibration by using weights 11 12 Rechargeable...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...applied is smaller than 105 the rated load after the meter is restarted Make sure the force gauge is not overloaded Overload may damage the sensor Do not use sharp tools to press buttons Keep the met...

Страница 5: ...for a test stand This reduces possible user error to a minimum and makes series of measurements possible 5 Characteristics Display LCD 180 rotatable backlit Interface USB Alarm Fracture alarm upper l...

Страница 6: ...PCE Instruments 4 7 Device description 7 1 Keys and adaptor 1 Measuring adaptor 2 Clamping nut 3 LCD 4 LED 5 Control panel 1 Switched output 2 USB interface 3 Power connection...

Страница 7: ...truments 5 7 2 Dimensions All dimensions in mm 7 3 Circuit diagram of 3 5 mm jack Here you can find the circuit diagram for connecting a 3 5 mm jack in order to establish a connection to the switched...

Страница 8: ...ighest value measured MIN lowest measured value 3 Displays current the peak value or current measurement value in real time depending on display orientation 4 Battery level indication 5 PC connection...

Страница 9: ...tion Zeroing Zeroing Memory is cleared x Reading and unit are sent x Data are sent Up Saved readings are qeried x Switch between AVER MAX and MIN Down Menu is opened Invalid Back to measurement mode S...

Страница 10: ...ndividually Change unit and leave a mode In test mode press to change the unit In the menu you can use to leave a selected option or menu In storage mode you can use to leave this mode and return to t...

Страница 11: ...ing saved values AVER MAX MIN In test mode you can open storage mode with the key In storage mode you can press to switch between AVER MAX and MIN USB interface Please observe the instructions in the...

Страница 12: ...arm deactivated outer outside alarm limit F 7 lo Lower alarm limit 0 F 8 hi Upper alarm limit Maximum F 9 cut 10 90 Overload alarm 50 F 10 peak key 3 60 sec Display time of peak value Remove by pressi...

Страница 13: ...attery This product is configured with a 1600 mAh 6 V Ni Hi rechargeable battery If it is fully charged the product can be running uninterruptedly for 10 hours When shut down the battery will discharg...

Страница 14: ...ecisive for the firmness of fruit i e the higher the temperature the softer the fruit will be 13 3 Important information about the measurement After preparing the sample and selecting the right tip yo...

Страница 15: ...ation tip 4 Peeled zone Try to make all measurements under equal conditions to ensure reproducible values for comparative and statistical purposes To achieve high accuracy use a test stand which enabl...

Страница 16: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Страница 17: ...583 975 114 Fax 49 0 2903 976 99 9971 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce instruments com info pce instruments co uk info pce italia it www pce instruments com deutsch www pce instrumen...

Отзывы: