Calibration procedure pH 10
1) Rinse the electrode in distilled water completely.
2) Connect the pH electrode to the meter and immerse it into the relevant calibration solution.
3) Power on the instrument.
4) Press the "CAL" key. The desired calibration value pH 7 will appear.
The display will show:
7.00
pH
CAL
5) Press the ▼ (Down) key until the desired calibration value appears:
10.00
pH
CAL
6) Press the "ENTER" key to confirm.
7
Important information on the calibration
- Before and after immersing the electrode into a pH solution, rinse it with distilled water.
- Calibrate each pH solution twice, if possible.
- It is recommended to calibrate in two steps: pH 7 first, followed by pH 4, then calibrate pH 7
again and then pH 10.
- If the calibration solution deviates by 2 pH,
“ERR” will appear in the display.