1
AH =
2,1668
P
ws
RH
100
(
T
+ 273,2)
gdzie:
T
→ measured temperature [
o
C]
RH
→ measured absolute humidity [%]
DP
→ dew-point temperature [
o
C]
P
ws
→ saturated water vapor pressure (water vapor pressure) [mbar]
AH
→ absolute humidity [g/m
3
]
DP
=
T
n
m
- 1
log
(
P
ws
RH
10000
A
)
Table 2
the factors used to calculate the dew-point
t [
o
C]
A
m
t
n
< 0
6.119866
7.926104
250.4138
0...50
6.1078
7.5
237.3
50...100
5.9987
7.3313
229.1
DP → dwe-point:
DP → absolute humidity:
5.5.1. Calculated values
Based on the measurement of a temperature and relative
humidity, the P18D transducer calculates dew-point and absolute humi-
dity of the following dependencies.
5.5.2. selection of controlling values for analog outputs
The P18(D) transducers in versions P18(D)-1XX, P18(D)-
2XX, P18(D)-4XX, P18(D)-5XX are equipped with two programmable
analog outputs (voltage or current). By default, the first analog output
is set for a presentation of the ambient temperature and the second
analog output is controlled by the value of the relative humidity. The ana-
log outputs can respond to any of the measuring or calculated values