Relative Mode 

The  relative  measurement  feature  allows  you  to  make  measurements  relative  to  a  stored 
reference  value.  A  reference  voltage,  current,  etc.  can  be  stored  and  measurements  made  in 
comparison to that value. The displayed value is the difference between the reference value and 
the measured value. Note: Relative mode does not operate in the 4-20mA function. 
1. Perform the measurement as described in the operating instructions. 
2. Press the REL button to store the reading in the display and the "REL" indicator will appear on 
    the display. 
3. Left auxiliary display displays the margin of initial value and the current value. Right auxiliary 
    display displays the initial reading. Main display the reading after REL TEST. 
4. Press the EXIT button to exit the relative mode. 

3.19 

Display Backlight 

Press the 

 key to turn the backlight on. The backlight will automatically turn off after SET time. 

Press the EXIT button to exit the backlight on mode. 

3.20 

HOLD 

The hold function freezes the reading in the display. Press the HOLD key momentarily to activate 
or to exit the HOLD function. 

3.21 

PEAK HOLD 

The Peak Hold function captures the peak AC or DC voltage or current. The meter can capture 
negative  or  positive  peaks  as  fast  as  1  millisecond  in  duration.  Momentarily  press  the  PEAK 
button, “PEAK” and “MAX” will display in left auxiliary display. MIN” will display in right auxiliary 
display.  The  meter  will  update  the  display  each  time  a  lower  negative  peak  occurs.  Press  the 
EXIT button to exit the PEAK HOLD mode. Auto Power Off feature will be disabled automatically 
in this mode. 

3.22 

Data Record (Store/Recall) 

1. STORE function 
In the current testing mode, press STORE button one time, enter into STORE function. On the 
left  upper  corner  of  LCD  shows  NO XXXX.,  which  states current storage  serial  number. Then, 
press button PEAKHOLD to change into the initial serial number 0000. (Press again it will change 
back). On the right upper corner of LCD shows XXXX, which states how many current storage is 
used. Press STORE button again, enter into recording interval time set up function. On the left 
upper shows 0000 S, which states recording interval time; using  & - to select, the range 
is 0~255 seconds. When the recording interval time is 0000 S, then press STORE button again 
to  change  into  manual  recording.  Press  the  STORE  button  again  to  record  once.  When  the 
recording interval time is 1~255 S, then press STORE button again to start recording automatically 
from 0000. Recording times is showed on the left upper corner, data is showed on the right upper 
corner  (Due  to  digital  limitation,  there  is  only  display  preceding  four  numbers)  To finish  above 
STORE function, press EXIT button shortly. If you expect to clear all memory data: While power 
on, hold EXIT button, and switch from OFF to random and then release EXIT button, the LCD will 
flash thrice and buzzer thrice too, which means all memory data is cleared. 
 

 

Содержание PCE-ITM 20

Страница 1: ...lation Meter PCE ITM 20 PCE ITM 20 User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last...

Страница 2: ...ents 10 3 7 Continuity Check 11 3 8 Diode Test 12 3 9 Capacitance Measurements 12 3 10 Temperature Measurements 13 3 11 Frequency Duty Cycle Measurements electronic 13 3 12 4 20mA Measurements 14 3 13...

Страница 3: ...PCE Instruments 4 Maintenance 19 4 1 Battery Installation 19 4 2 Replacing the Fuses 20 5 Technical specifications 20 6 Test method of DAR and PI 24 7 Warranty 25 8 Disposal 25...

Страница 4: ......

Страница 5: ...he specifications must not be exceeded under any circumstances NEVER apply voltage or current to the meter that exceeds the specified maximum Input Protection Limits Function Maximum input V DC or V A...

Страница 6: ...round exceeds 1000 VAC or VDC This symbol adjacent to one or more terminals identifies them as being associated with ranges that may in normal use be subject to particularly hazardous voltages For max...

Страница 7: ...ETUP button 5 MODE LOCK button 6 INSULATION TEST button 7 Function switch 8 10 A input jacks 9 mA A and insulation input jacks 10 COM input jack 11 Positive input jack 12 Backlight button 13 EXIT AC D...

Страница 8: ...3 volts amps Amps kilo 103 ohms Farads capacitance mega 106 ohms Ohms Peak Hold Hertz frequency Volts Per cent duty ratio Relative Alternating current Autoranging Direct current Display hold Degrees F...

Страница 9: ...ation Risk of electrocution High voltage circuits both AC and DC are very dangerous and should be measured with great care ALWAYS turn the function switch to the OFF position when the meter is not in...

Страница 10: ...ur that can damage the meter Set the function switch to the green VAC Hz position Insert the black test lead banana plug into the negative COM jack Insert red test lead banana plug into the positive V...

Страница 11: ...ate DC or AC or in AC range press EXIT for two seconds and choose AC DC Insert the black test lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive V jack...

Страница 12: ...500 mA DC set the function switch to the yellow mA position and insert the red test lead banana plug into the A mA jack For current measurements up to 20 A DC set the function switch to the yellow 10...

Страница 13: ...he red test lead banana plug into the 10A jack Press the MODE button to indicate AC on the display Remove power from the circuit under test then open up the circuit at the point where you wish to meas...

Страница 14: ...tion switch to the green CAP position Insert the black test lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive jack Press the MODE button to indicate on...

Страница 15: ...osition Insert the black lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive jack Press the MODE button to indicate and on the display Touch the test pro...

Страница 16: ...ices will indicate near 0V and an open device will indicate OL in both polarities 3 9 Capacitance Measurements To avoid electric shock disconnect power to the unit under test and discharge all capacit...

Страница 17: ...t 30 seconds Read the temperature in the display Note The temperature probe is fitted with a K type mini connector A mini connector to banana connector adaptor is supplied for connection to the input...

Страница 18: ...ll be damaged by insulation measurement voltage Many parts will be damaged undue 1000V and other ranges For example power factor capacitor low voltage insulation cable electronic dimmer and common ele...

Страница 19: ...tor wire and all other components between the motor and the main switch If a weakness is indicated the motor and other components should be checked individually If the motor is disconnected at the mot...

Страница 20: ...e best mode for most measurements For measurement situations requiring that a range be manually selected perform the following 1 Press the RANGE key The AUTO display indicator will turn off 2 Press th...

Страница 21: ...lay MIN will display in right auxiliary display The meter will update the display each time a lower negative peak occurs Press the EXIT button to exit the PEAK HOLD mode Auto Power Off feature will be...

Страница 22: ...t includes in sequence A upper limit buzzer alarm B lower limit buzzer alarm C auto power off time D turn off phonating E back lit time Use buttons to select the parameter 2 Press SET button continuou...

Страница 23: ...O NOT use chemicals cleaning solvents or detergents USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE Remove old or weak batteries so they do not leak and damage the unit IF THE METER IS TO BE...

Страница 24: ...meter until the fuse cover is in place and fastened securely 5 Technical specifications Test voltage test current Measuring range Resolution Accuracy 50V 1 mA 50 k 0 01 5 M 0 0001 M 4 20 digits 5 50...

Страница 25: ...gits 500 V 0 01 V 1 30 digits 1000 V 0 1 V 1 30 digits The specified accuracy for alternating voltage refers to 5 100 of the measuring range Direct current Measuring range Resolution Accuracy 500 A 0...

Страница 26: ...digits 500 Hz 0 01 Hz 0 02 3 digits 5 kHz 0 0001 kHz 0 02 3 digits 50 kHz 0 001 kHz 0 02 3 digits 500 kHz 0 01 kHz 0 02 3 digits 5 MHz 0 0001 MHz 0 02 3 digits 50 MHz 0 001 MHz 0 02 3 digits 100 MHz...

Страница 27: ...at 500V 1 5 at 1000V Display 50000 digit LCD illuminated with bargraph Automatic shutdown after 15 minutes Measurement rate update display 10 Hz Power supply 6 x 1 5V AA batteries Interface Wireless U...

Страница 28: ...cally 3 1 1 5 1 1 0 or less Tested material status Considered as good insulation older types Not acceptable older types very high insulation resistance Modern type of good insulation system Fail Examp...

Страница 29: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Страница 30: ...dustrial needs com info pce italia it www pce instruments com deutsch www pce instruments com english www pce instruments com italiano The Netherlands Chile Hong Kong PCE Brookhuis B V PCE Instruments...

Отзывы: