3 4
Safety symbols:
Caution refer to this manual before using the meter.
Dangerous voltages.
Meter is protected throughout by double insulation or
reinforced insulation.
When servicing, use only specified replacement parts.
CE
Comply with EN-61010-1
1. SPECIFICATIONS
1-1 General Information
Environment conditions:
①
①
①
①
Installation Categories II
②
②
②
②
Pollution Degree 2
③
③
③
③
Altitude up to 2000 meters
④
④
④
④
Indoor use only
⑤
⑤
⑤
⑤
Relatively humidity 80% max.
⑥
⑥
⑥
⑥
Operation Ambient 0~40ºC
Maintenance & Clearing:
①
①
①
①
Repairs or servicing not covered in this manual should
only be performed by qualified personnel.
②
②
②
②
Periodically wipe the case with a dry cloth. Do not use
abrasives or solvents on this instruments.
Display:
Large LCD with dual display
Measurement Range:
200
Ω
, 200k
Ω
, 200M
Ω
/250V,
200M
Ω
/500V, 2000M
Ω
/1000V,
750V/ACV, 1000V/DCV.
Sampling Rate:
2.5 times per second.
Zero Adjustment:
Automatic adjustment.
Over Range Indicator:
Number 1 of highest digit is
displayed.
Low Battery Indication:
The is displayed when the
battery Voltage drop below the
operating voltage.
Operating Temperature:
0ºC to 40ºC (32ºF to 104ºF) and
Humidity below 80% RH
Storage Temperature:
-10ºC to 60ºC (14ºF to 140ºF) and
Humidity below 70% RH
Power source:
DC9V (6x1.5V Size “AA” battery or Equivalent)
Dimensions:
200(L) x 92(W) x 50(H) mm
Weight:
Approx 700g include battery
Accessories:
Test leads, 6pcs battery, Carrying case, manual.