ZERO CALIBRATION

Zero  adjustment  for  'Fe'  and  'NFe'    should  be

 

carried

out separately. Take the iron substrate if

 

'Fe'  on

Display,  while  take  the  aluminium

 

substrate if 'NFe'

on the Display. Place the probe

 

(3-1)  on  the  substrate

steadily.    Press  the  Zero

 

key  (3-3)  and    '0'    will    be

on    the  Display

 

before lifting the probe. 

If pressing

the ZERO

 

key  but  the  probe  is  not  placed  on

the

 

substrate  or  an  uncoated  standard.  The  zero

calibration  is invalid.

battery  voltage  less  than  approx. 

2

.5v  ,  the 

battery

 

symbol

 

'

               

'  will  appear  on  the 

Display 

5

6.2 Slide  the  Battery Cover  ( Fig. 1 ,  3-4) away 

from the instrument and remove the batteries. 

6.3  Install  the  batteries  (

2

x1.5v  AAA/UM-4) 

correctly into the case.  

6.4  If    the    instrument  is  not  to  be  used    for  any 

extended period , remove batteries.   

6. BATTERY REPLACEMENT

6.1 When it is necessary to replace the battery, i.e.

6

7.CALIBRATION FOILS

A s  a c c e s s o r i e s ,  t h e  i n s t r u m e n t  i n c l u d e s  a

different  foil  set  for  different  ranges.  Please  see
the following table.

STANDARD FOIL INCLUDED 

RANGE 

(um) 

CM25 

CM50 

CM100 

CM200 

CM500 

CM1000 

0~200 

0~500 

0~1000 

0~2000 

Customized 

7

8. CONSIDERATIONS

8.1  In order to weaken the influence of the measured

material  on  the  accuracy  of  measurement,  it  is 
recommended  that  the  calibrations  should  be 
done on the uncoated material to be measured.

8.2  Probes  will  eventually  wear.  Probe  life  will 

depend on the number of measurements taken and 
how  abrasive  the  coating  is.  Replacement 
separate can be fitted by qualified persons only.